DICOM PS3.18 2019a - Web Services

6.10.3.2 Retrieve Transaction

The Retrieve transaction retrieves the target NPI resource in a DICOM Media Type.

6.10.3.2.1 Request

The Retrieve request has the following syntax:

    GETSP /{npi-name}/{uid} SP version CRLF
    Accept: 1#dicom-media-type CRLF
    [If-None-Match: entity-tag CRLF]
    *(header-field CRLF)
    CRLF
6.10.3.2.1.1 Resources

The target URL shall reference one of the resources shown in Table 6.10.3.2.1.1-1.

An origin server shall specify all supported npi-names in its Conformance Statement and in its response to the Retrieve Capabilities transaction.

Table 6.10.3.2.1.1-1. Resources and URI Templates

Resource

URI Template

Instance

/{npi-name}/{uid}


6.10.3.2.1.2 Query Parameters

There are no additional query parameters.

6.10.3.2.1.3 Request Header Fields

Table 6.10.3.2.1.3-1 shows the most common Mandatory, Conditional, and common Optional header fields for this transaction.

Table 6.10.3.2.1.3-1. Request Header Fields

Header Field

Value

Usage

Accept

dicom-media-type

M


6.10.3.2.1.4 Request Payload

The request shall have no payload.

DICOM PS3.18 2019a - Web Services