DICOM PS3.17 2020a - Explanatory Information

XXX.3.7.3 Encoding Details

XXX.3.7.3.1 Volumetric Presentation State Relationship Module Recommendations

Table XXX.3.7-1. Volumetric Presentation State Relationship Module Recommendations

Attribute Name

Tag

Comment

Volumetric Presentation State Input Sequence

(0070,1201)

Two items are this sequence referencing one volume for each data type

>ITEM 1

>Volumetric Presentation Input Number

(0070,1207)

Set to 1

>Presentation Input Type

(0070,1202)

Set to "VOLUME".

>Referenced Image Sequence

(0008,1140)

Sequence of frames with Data Type (0018,9808) value of TISSUE_INTENSITY

>>Table 10-3 “Image SOP Instance Reference Macro Attributes” in PS3.3

>Crop

(0070,1204)

Set to "NO".

>ITEM 2

>Volumetric Presentation Input Number

(0070,1207)

Set to 2

>Presentation Input Type

(0070,1202)

Set to "VOLUME".

>Referenced Image Sequence

(0008,1140)

Sequence of frames with Data Type (0018,9808) value of FLOW_VELOCITY

>>Table 10-3 “Image SOP Instance Reference Macro Attributes” in PS3.3

>Crop

(0070,1204)

Set to "NO".


DICOM PS3.17 2020a - Explanatory Information