DICOM PS3.17 2019b - Explanatory Information

V.5 Hanging Protocol Query Example

The following is an example of a general C-FIND Request for the Hanging Protocol Information Model - FIND SOP Class that is searching for all Chest related Hanging Protocols for the purpose of reading projection Chest X-ray. The user is at a workstation that has two 2Kx2.5K screens.

C-FIND Request:

Nesting

Attribute

Tag

VR

VL (hex)

Value

Affected SOP Class UID

(0000,0002)

UI

0018

1.2.840.10008.5.1.4.38.2

Command Field

(0000,0100)

US

0002

0020H [C-FIND-RQ]

Message ID

(0000,0110)

US

0002

0010H

Priority

(0000,0700)

US

0002

0000H [MEDIUM]

Data Set Type

(0000,0800)

US

0002

0102H

SOP Class UID

(0008,0016)

UI

0000

SOP Instance UID

(0008,0018)

UI

0000

Hanging Protocol Name

(0072,0002)

SH

0000

Hanging Protocol Description

(0072,0004)

LO

0000

Hanging Protocol Level

(0072,0006)

CS

0000

Hanging Protocol Creator

(0072,0008)

LO

0000

Hanging Protocol Creation DateTime

(0072,000A)

DT

0000

Hanging Protocol Definition Sequence

(0072,000C)

SQ

ffffffff

%item

>

Modality

(0008,0060)

CS

0000

>

Anatomic Region Sequence

(0008,2218)

SQ

ffffffff

%item

>>

Code Value

(0008,0100)

SH

0008

51185008

>>

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

>>

Code Meaning

(0008,0104)

LO

0006

Chest

%enditem

%endseq

>

Procedure Code Sequence

(0008,1032)

SQ

0000

>

Laterality

(0020,0060)

CS

0000

>

Reason for Requested Procedure Code Sequence

(0040,100A)

SQ

0000

%enditem

%endseq

Hanging Protocol User Identification Code Sequence

(0072,000E)

SQ

0000

Number of Priors Referenced

(0072,0014)

US

0000

Number of Screens

(0072,0100)

US

0000

Nominal Screen Definition Sequence

(0072,0102)

SQ

0000

The following is an example of a set of C-FIND Responses for the Hanging Protocol Information Model - FIND SOP Class, answering the C-FIND Request listed above. There are a few matches for this general query. The application needs to select the best choice among the matches, which is the second response. The first response is for Chest CT, and the third response does not match the user's workstation environment as well as does the second.

C-FIND Response #1:

Nesting

Attribute

Tag

VR

VL (hex)

Value

Affected SOP Class UID

(0000,0002)

UI

0018

1.2.840.10008.5.1.4.38.2

Command Field

(0000,0100)

US

0002

8020H [C-FIND-RSP]

Message ID Being Responded To

(0000,0120)

US

0002

0010H

Data Set Type

(0000,0800)

US

0002

0102H

Status

(0000,0900)

US

0002

FF00H [Pending]

SOP Class UID

(0008,0016)

UI

0018

1.2.840.10008.5.1.4.38.1

SOP Instance UID

(0008,0018)

UI

0024

1.2.840.10008.5.1.4.1.1.76392.999.2

Hanging Protocol Name

(0072,0002)

SH

000a

CT 1 prior

Hanging Protocol Description

(0072,0004)

LO

0038

Dual screen layout for current and single prior chest CT

Hanging Protocol Level

(0072,0006)

CS

000c

SINGLE_USER

Hanging Protocol Creator

(0072,0008)

LO

0008

Dr. Chan

Hanging Protocol Creation DateTime

(0072,000A)

DT

000c

200408210718

Hanging Protocol Definition Sequence

(0072,000C)

SQ

ffffffff

%item

>

Modality

(0008,0060)

CS

0002

CT

>

Anatomic Region Sequence

(0008,2218)

SQ

ffffffff

%item

>>

Code Value

(0008,0100)

SH

0008

51185008

>>

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

>>

Code Meaning

(0008,0104)

LO

0006

Chest

%enditem

%endseq

>

Procedure Code Sequence

(0008,1032)

SQ

0000

>

Laterality

(0020,0060)

CS

0000

>

Reason for Requested Procedure Code Sequence

(0040,100A)

SQ

0000

%enditem

%endseq

Hanging Protocol User Identification Code Sequence

(0072,000E)

SQ

0000

%item

>

Code Value

(0008,0100)

SH

000a

58489749P

>

Coding Scheme Designator

(0008,0102)

SH

0008

HOSP_ID

>

Code Meaning

(0008,0104)

LO

000e

Susan H. Chan

%enditem

%endseq

Number of Priors Referenced

(0072,0014)

US

0002

1

Number of Screens

(0072,0100)

US

0002

2

Nominal Screen Definition Sequence

(0072,0102)

SQ

0000

C-FIND Response #2:

Nesting

Attribute

Tag

VR

VL (hex)

Value

Affected SOP Class UID

(0000,0002)

UI

0018

1.2.840.10008.5.1.4.38.2

Command Field

(0000,0100)

US

0002

8020H [C-FIND-RSP]

Message ID Being Responded To

(0000,0120)

US

0002

0010H

Data Set Type

(0000,0800)

US

0002

0102H

Status

(0000,0900)

US

0002

FF00H [Pending]

SOP Class UID

(0008,0016)

UI

0018

1.2.840.10008.5.1.4.38.1

SOP Instance UID

(0008,0018)

UI

0020

1.2.840.123456.20030822.223344.1

Hanging Protocol Name

(0072,0002)

SH

000c

Chest X-ray

Hanging Protocol Description

(0072,0004)

LO

0026

Current and Prior Chest PA and Lateral

Hanging Protocol Level

(0072,0006)

CS

0004

SITE

Hanging Protocol Creator

(0072,0008)

LO

0012

Senior Radiologist

Hanging Protocol Creation DateTime

(0072,000A)

DT

000e

20020823133455

Hanging Protocol Definition Sequence

(0072,000C)

SQ

ffffffff

%item

>

Modality

(0008,0060)

CS

0000

>

Anatomic Region Sequence

(0008,2218)

SQ

ffffffff

%item

>>

Code Value

(0008,0100)

SH

0008

51185008

>>

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

>>

Code Meaning

(0008,0104)

LO

0006

Chest

%enditem

%endseq

>

Procedure Code Sequence

(0008,1032)

SQ

0000

>

Laterality

(0020,0060)

CS

0000

>

Reason for Requested Procedure Code Sequence

(0040,100A)

SQ

0000

%enditem

%endseq

Hanging Protocol User Identification Code Sequence

(0072,000E)

SQ

0000

Number of Priors Referenced

(0072,0014)

US

0002

1

Number of Screens

(0072,0100)

US

0002

0002H

Nominal Screen Definition Sequence

(0072,0102)

SQ

ffffffff

%item

>

Number of Vertical Pixels

(0072,0104)

US

0002

2560

>

Number of Horizontal Pixels

(0072,0106)

US

0002

2048

>

Display Environment Spatial Position

(0072,0108)

FD

0020

0.0\1.0\0.5\0.0

>

Screen Minimum Grayscale Bit Depth

(0072,010A)

US

0002

0008H

>

Application Maximum Repaint Time

(0072,010E)

US

0002

0064H

%enditem

%item

>

Number of Vertical Pixels

(0072,0104)

US

0002

2560

>

Number of Horizontal Pixels

(0072,0106)

US

0002

2048

>

Display Environment Spatial Position

(0072,0108)

FD

0020

0.5\1.0\1.0\0.0

>

Screen Minimum Grayscale Bit Depth

(0072,010A)

US

0002

0008H

>

Application Maximum Repaint Time

(0072,010E)

US

0004

0064H

%enditem

%endseq

C-FIND Response #3:

Nesting

Attribute

Tag

VR

VL (hex)

Value

Affected SOP Class UID

(0000,0002)

UI

0018

1.2.840.10008.5.1.4.38.2

Command Field

(0000,0100)

US

0002

8020H [C-FIND-RSP]

Message ID Being Responded To

(0000,0120)

US

0002

0010H

Data Set Type

(0000,0800)

US

0002

0102H

Status

(0000,0900)

US

0002

FF00H [Pending]

SOP Class UID

(0008,0016)

UI

0018

1.2.840.10008.5.1.4.38.1

SOP Instance UID

(0008,0018)

UI

002a

1.2.840.113986.2.664566.21121125.85669.967

Hanging Protocol Name

(0072,0002)

SH

0010

Chest X-ray_LGon

Hanging Protocol Description

(0072,0004)

LO

003e

Prior and Current Lateral of Chest X-ray for two screen system

Hanging Protocol Level

(0072,0006)

CS

000c

SINGLE_USER

Hanging Protocol Creator

(0072,0008)

LO

0012

Dr. Leia Gonzales

Hanging Protocol Creation DateTime

(0072,000A)

DT

000e

20030822101100

Hanging Protocol Definition Sequence

(0072,000C)

SQ

ffffffff

%item

>

Modality

(0008,0060)

CS

0002

DX

>

Anatomic Region Sequence

(0008,2218)

SQ

ffffffff

%item

>>

Code Value

(0008,0100)

SH

0008

51185008

>>

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

>>

Code Meaning

(0008,0104)

LO

0006

Chest

%enditem

%endseq

>

Procedure Code Sequence

(0008,1032)

SQ

0000

>

Laterality

(0020,0060)

CS

0000

>

Reason for Requested Procedure Code Sequence

(0040,100A)

SQ

0000

%enditem

%endseq

Hanging Protocol User Identification Code Sequence

(0072,000E)

SQ

0000

%item

>

Code Value

(0008,0100)

SH

0004

Lgon

>

Coding Scheme Designator

(0008,0102)

SH

0008

99Local

>

Coding Scheme Version

(0008,0103)

SH

0004

v40a

>

Code Meaning

(0008,0104)

LO

000c

log-in name

%enditem

%endseq

Number of Priors Referenced

(0072,0014)

US

0002

1

Number of Screens

(0072,0100)

US

0002

0002H

Nominal Screen Definition Sequence

(0072,0102)

SQ

ffffffff

%item

>

Number of Vertical Pixels

(0072,0104)

US

0002

1280

>

Number of Horizontal Pixels

(0072,0106)

US

0002

1024

>

Display Environment Spatial Position

(0072,0108)

FD

0020

0.0\1.0\0.5\0.0

>

Screen Minimum Grayscale Bit Depth

(0072,010A)

US

0002

0008H

>

Application Maximum Repaint Time

(0072,010E)

US

0004

0064H

%enditem

%item

>

Number of Vertical Pixels

(0072,0104)

US

0002

1280

>

Number of Horizontal Pixels

(0072,0106)

US

0002

1024

>

Display Environment Spatial Position

(0072,0108)

FD

0020

0.5\1.0\1.0\0.0

>

Screen Minimum Grayscale Bit Depth

(0072,010A)

US

0002

0008H

>

Application Maximum Repaint Time

(0072,010E)

US

0004

0064H

%enditem

%endseq

C-FIND Response #4:

Nesting

Attribute

Tag

VR

VL (hex)

Value

Affected SOP Class UID

(0000,0002)

UI

0018

1.2.840.10008.5.1.4.38.2.

Command Field

(0000,0100)

US

0002

8020H [C-FIND-RSP]

Message ID Being Responded To

(0000,0120)

US

0002

0010H

Data Set Type

(0000,0800)

US

0002

0101H

Status

(0000,0900)

US

0002

0000H [Success]

DICOM PS3.17 2019b - Explanatory Information