DICOM PS3.17 2019a - Explanatory Information

V.4 Neurosurgery Planning Hanging Protocol Example

Goal: A Hanging Protocol for MR & CT of Head, for a neurosurgery plan. 1Kx1K screen on left shows orthogonal MPR slices through the acquisition volume, and in one presentation group has a 3D interactive volume rendering in the lower right quadrant. In all display sets the 1Kx1K screen is split into 4 512x512 quadrants. The 2560x2048 screen has a 4 row by 3 column tiled display area. There are 4 temporal presentation groups: CTnew, MR, combined CTnew and MR, combined CTnew and CTold.

Display Environment Spatial Position Attribute values for image boxes are represented in terms of ratios in pixel space [(0/3072, 512/2560), (512/3072,0/2560)] rather than (0.0,0.0), (1.0,1.0) space, for ease of understanding the example.

Neurosurgery Planning Hanging Protocol Example

Figure V.4-1. Neurosurgery Planning Hanging Protocol Example


V.4.1 Hanging Protocol Definition Module

  • Hanging Protocol Name: "NeurosurgeryPlan"

  • Hanging Protocol Description: "Neurosurgery planning, requiring MR and CT of head"

  • Hanging Protocol Level: "SITE"

  • Hanging Protocol Creator: "Smith^Joseph"

  • Hanging Protocol Creation DateTime: "20020101104200"

  • Hanging Protocol Definition Sequence:

    • Item 1:

      • Modality: "MR"

      • Anatomic Region Sequence:

      • Laterality: zero length

      • Procedure Code Sequence:

        • Item 1: (98765, 99Local, 1.5, "NeuroSurgery Plan Local5")

      • Reason for Requested Procedure Code Sequence:

        • Item 1: (I67.1, I10, "Cerebral aneurysm")

    • Item 2:

      • Modality: "CT"

      • Anatomic Region Sequence:

      • Laterality: zero length

      • Procedure Code Sequence:

        • Item 1: (98765, 99Local, 1.5, "NeuroSurgery Plan Local5")

      • Reason for Requested Procedure Code Sequence:

        • Item 1: (I67.1, I10, "Cerebral aneurysm")

  • Number of Priors Referenced: 1

  • Image Sets Sequence:

    • Item 1:

      • Image Set Selector Sequence:

        • o Item 1:

          • Image Set Selector Usage Flag: "NO_MATCH"

          • Selector Attribute: (0018,0015) [Body Part Examined]

          • Selector Attribute VR: "CS"

          • Selector CS Value: "HEAD"

          • Selector Value Number: 1

        • Item 2:

          • Image Set Selector Usage Flag: "NO_MATCH"

          • Selector Attribute: (0008,0060) [Modality]

          • Selector Attribute VR: "CS"

          • Selector CS Value: "MR"

          • Selector Value Number: 1

      • Time Based Image Sets Sequence:

        • o Item 1:

          • Image Set Number: 1

          • Image Set Selector Category: "RELATIVE_TIME"

          • Relative Time: 0\0

          • Relative Time Units: "MINUTES"

          • Image Set Label: "Current MR Head"

    • Item 2:

      • Image Set Selector Sequence:

        • Item 1:

          • Image Set Selector Usage Flag: "NO_MATCH"

          • Selector Attribute: (0018,0015) [Body Part Examined]

          • Selector Attribute VR: "CS"

          • Selector CS Value: "HEAD"

          • Selector Value Number: 1

        • o Item 2:

          • Image Set Selector Usage Flag: "NO_MATCH"

          • Selector Attribute: (0008,0060) [Modality]

          • Selector Attribute VR: "CS"

          • Selector CS Value: "CT"

          • Selector Value Number: 1

      • Time Based Image Sets Sequence:

        • Item 1:

          • Image Set Number: 2

          • Image Set Selector Category: "RELATIVE_TIME"

          • Relative Time: 0\0

          • Relative Time Units: "MINUTES"

          • Image Set Label: "Current CT Head"

        • Item 2:

          • Image Set Number: 3

          • Image Set Selector Category: "ABSTRACT_PRIOR"

          • Abstract Prior Value: 1\1

          • Image Set Label: "Prior CT Head"

  • Hanging Protocol User Identification Code Sequence: zero length

  • Hanging Protocol User Group Name: "ABC Hospital"

DICOM PS3.17 2019a - Explanatory Information