DICOM PS3.17 2019b - Explanatory Information

PPP.3.2 An Example of A Tablet Display

A Tablet Display System is shown in Figure PPP.3.2-1.

A Tablet Display System

Figure PPP.3.2-1. A Tablet Display System


The following is an example of an N-GET Request/Response pair for the Display System SOP Class.

This example is encoded with Undefined Sequence Length and Undefined Item Length, so it contains Sequence Delimitation Items and Item Delimitation Items.

N-GET:

ANP

Attribute Not Present.

VNP

Attribute Present but Value Not Present.

-

Not specified.

Table PPP.3.2-1. N-GET Request/Response Example

Attribute Name

Tag

N-GET Request (SCU)

N-GET Response (SCP)

SOP Common and Workstation Modules

Specific Character Set

(0008,0005)

ANP

\ISO 2022 IR 87

Manufacturer

(0008,0070)

VNP

Tablet Corp.

Institution Name

(0008,0080)

VNP

JIRA Hospital

Institution Address

(0008,0081)

VNP

Bunkyo-ku, Tokyo, Japan

Device Serial Number

(0018,1000)

VNP

AA1B22CCCC3D

Station Name

(0008,1010)

VNP

TABLET1

Institutional Department Name

(0008,1040)

VNP

Radiology Dept.

Manufacturer's Model Name

(0008,1090)

VNP

MC706J/A

Equipment Administrator Sequence

(0028,7000)

VNP

>Item #1 of Equipment Administrator Sequence

(FFFE,E000)

-

>Person Name

(0040,A123)

-

Yamada^Tarou=山田^太郎=やまだ^たろう

>Person Identification Code Sequence

(0040,1101)

-

>>Item #1 of Person Identification Code Sequence

(FFFE,E000)

-

>>Code Value

(0008,0110)

-

111111

>>Coding Scheme Designator

(0008,0102)

-

LOCAL

>>Code Meaning

(0008,0104)

-

Yamada^Tarou

>>Item Delimiter of Item #1 of Person Identification Code Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Person Identification Code Sequence

(FFFE,E0DD)

-

>Person's Address

(0040,1102)

-

>Person's Telephone Number

(0040,1103)

-

EXT. 1234

>Institution Name

(0008,0080)

-

IT Support Div.

>Item Delimiter of Item #1 of Equipment Administrator Sequence

(FFFE,E00D)

-

>Sequence Delimiter of Equipment Administrator Sequence

(FFFE,E0DD)

-

Display System Module

Number of Display Subsystems

(0028,7001)

VNP

1

Display Subsystem Sequence

(0028,7023)

VNP

Item #1 of Display Subsystem Sequence

(FFFE,E000)

-

>Display Subsystem ID

(0028,7003)

-

1

>Display Subsystem Name

(0028,7004)

-

DS1

>Display Subsystem Description

(0028,7005)

-

Embedded LCD

>Display Device Type Code Sequence

(0028,7022)

-

>>Item #1 of Display Device Type Code Sequence

(FFFE,E000)

-

>>Code Value

(0008,0100)

-

109992

>>Coding Scheme Designator

(0008,0102)

-

DCM

>>Code Meaning

(0008,0104)

-

Liquid Crystal Display

>>Item Delimiter of Item #1 of Display Device Type Code Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Display Device Type Code Sequence

(FFFE,E0DD)

-

>Manufacturer

(0008,0070)

-

Tablet Corp.

>Device Serial Number

(0018,1000)

-

AA1B22CCCC3D

>Manufacturer's Model Name

(0008,1090)

-

MC706J/A

>System Status

(0028,7006)

-

NORMAL

>System Status Comment

(0028,7007)

-

>Display Subsystem Configuration Sequence

(0028,700A)

-

>>Item #1 of Display Subsystem Configuration Sequence

(FFFE,E000)

-

>>Configuration ID

(0028,700B)

-

1

>>Configuration Name

(0028,700C)

-

DS1Config1

>> Configuration Description

(0028,700D)

-

Configuration1 of Display Subsystem ID1

>>Referenced Target Luminance Characteristics ID

(0028,700E)

-

1

>>Item Delimiter of Item #1 of Display Subsystem Configuration Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Display Subsystem Configuration Sequence

(FFFE,E0DD)

-

>Current Configuration ID

(0028,7002)

-

1

>Measurement Equipment Sequence

(0028,7012)

-

>Sequence Delimiter of Measurement Equipment Sequence

(FFFE,E0DD)

-

>Item Delimiter of Item #1 of Display System Sequence

(FFFE,E00D)

-

>Sequence Delimiter of Display Subsystem Sequence

(FFFE,E0DD)

-

Target Luminance Characteristics Module

Target Luminance Characteristics Sequence

(0028,7008)

VNP

>Item #1 of Target Luminance Characteristics Sequence

(FFFE,E000)

-

>Luminance Characteristics ID

(0028,7009)

-

1

>Display Function Type

(0028,7019)

-

GAMMA

>Target Minimum Luminance

(0028,701D)

-

0.75

>Target Maximum Luminance

(0028,701E)

-

300

>Gamma Value

(0028,701A)

-

2.2

>Item Delimiter of Item #1 of Target Luminance Characteristics Sequence

(FFFE,E00D)

-

>Sequence Delimiter of Target Luminance Characteristics Sequence

(FFFE,E0DD)

-

QA Result Module

QA Results Sequence

(0028,700F)

VNP

>Item #1 of QA Results Sequence

(FFFE,E000)

-

>Display Subsystem ID

(0028,7003)

-

1

>Display Subsystem QA Results Sequence

(0028,7010)

-

There are no items in this sequence in this example.


DICOM PS3.17 2019b - Explanatory Information