DICOM PS3.17 2019b - Explanatory Information

E.3.2 Example 2: Calcification and Mass Detection With Findings

A mammography CAD device processes a screening mammography case with four films and a mass in the left breast. Mammography CAD runs both density and calcification detection successfully. It finds two densities in the LCC, one density in the LMLO, a cluster of two calcifications in the RCC and a cluster of 20 calcifications in the RMLO. It performs two clustering algorithms. One identifies individual calcifications and then clusters them, and the second simply detects calcification clusters. It performs mass correlation and combines one of the LCC densities and the LMLO density into a mass; the other LCC density is flagged Not for Presentation, therefore not intended for display to the end-user. The mammograms resemble:

Mammograms as Described in Example 2

Figure E.3-2. Mammograms as Described in Example 2


The content tree structure in this example is complex. Structural illustrations of portions of the content tree are placed within the content tree table to show the relationships of data within the tree. Some content items are duplicated (and shown in boldface) to facilitate use of the diagrams.

Content Tree Root of Example 2 Content Tree

Figure E.3-3. Content Tree Root of Example 2 Content Tree


Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1

Mammography CAD Report

TID 4000

1.1

Image Library

TID 4000

1.2

CAD Processing and Findings Summary

All algorithms succeeded; with findings

TID 4001

1.3

Summary of Detections

Succeeded

TID 4000

1.4

Summary of Analyses

Succeeded

TID 4000

Image Library Branch of Example 2 Content Tree

Figure E.3-4. Image Library Branch of Example 2 Content Tree


Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1.1

Image Library

TID 4000

1.1.1

IMAGE 1

TID 4020

1.1.1.1

Image Laterality

Right

TID 4020

1.1.1.2

Image View

Cranio-caudal

TID 4020

1.1.1.3

Study Date

19990101

TID 4020

1.1.2

IMAGE 2

TID 4020

1.1.2.1

Image Laterality

Left

TID 4020

1.1.2.2

Image View

Cranio-caudal

TID 4020

1.1.2.3

Study Date

19990101

TID 4020

1.1.3

IMAGE 3

TID 4020

1.1.3.1

Image Laterality

Right

TID 4020

1.1.3.2

Image View

Medio-lateral oblique

TID 4020

1.1.3.3

Study Date

19990101

TID 4020

1.1.4

IMAGE 4

TID 4020

1.1.4.1

Image Laterality

Left

TID 4020

1.1.4.2

Image View

Medio-lateral oblique

TID 4020

1.1.4.3

Study Date

19990101

TID 4020

CAD Processing and Findings Summary Bifurcation of Example 2 Content Tree

Figure E.3-5. CAD Processing and Findings Summary Bifurcation of Example 2 Content Tree


Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1.2

CAD Processing and Findings Summary

All algorithms succeeded; with findings

TID 4001

1.2.1

Individual Impression/Recommendation

TID 4003

1.2.2

Individual Impression/Recommendation

TID 4003

1.2.3

Individual Impression/Recommendation

TID 4003

1.2.4

Individual Impression/Recommendation

TID 4003

Individual Impression/Recommendation 1.2.1 from Example 2 Content Tree

Figure E.3-6. Individual Impression/Recommendation 1.2.1 from Example 2 Content Tree


Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1.2.1

Individual Impression/Recommendation

TID 4003

1.2.1.1

Rendering Intent

Presentation Required

TID 4003

1.2.1.2

Composite Feature

Mass

TID 4004

1.2.1.2.1

Rendering Intent

Presentation Required

TID 4004

1.2.1.2.2

Composite type

Target content items are related spatially

TID 4005

1.2.1.2.3

Scope of Feature

Feature was detected on multiple images

TID 4005

1.2.1.2.4

Algorithm Name

"Mass Maker"

TID 4019

1.2.1.2.5

Algorithm Version

"V1.9"

TID 4019

1.2.1.2.6

Single Image Finding

Mammography breast density

TID 4006

1.2.1.2.7

Single Image Finding

Mammography breast density

TID 4006

Single Image Finding Density 1.2.1.2.6 from Example 2 Content Tree

Figure E.3-7. Single Image Finding Density 1.2.1.2.6 from Example 2 Content Tree


Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1.2.1.2.6

Single Image Finding

Mammography breast density

TID 4006

1.2.1.2.6.1

Rendering Intent

Presentation Required

TID 4006

1.2.1.2.6.2

Algorithm Name

"Density Detector"

TID 4019

1.2.1.2.6.3

Algorithm Version

"V3.7"

TID 4019

1.2.1.2.6.4

Center

POINT

TID 4021

1.2.1.2.6.4.1

Reference to node 1.1.2

TID 4021

1.2.1.2.6.5

Outline

SCOORD

TID 4021

1.2.1.2.6.5.1

Reference to node 1.1.2

TID 4021

Single Image Finding Density 1.2.1.2.7 from Example 2 Content Tree

Figure E.3-8. Single Image Finding Density 1.2.1.2.7 from Example 2 Content Tree


Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1.2.1.2.7

Single Image Finding

Mammography breast density

TID 4006

1.2.1.2.7.1

Rendering Intent

Presentation Required

TID 4006

1.2.1.2.7.2

Algorithm Name

"Density Detector"

TID 4019

1.2.1.2.7.3

Algorithm Version

"V3.7"

TID 4019

1.2.1.2.7.4

Center

POINT

TID 4021

1.2.1.2.7.4.1

Reference to node 1.1.4

TID 4021

1.2.1.2.7.5

Outline

SCOORD

TID 4021

1.2.1.2.7.5.1

Reference to node 1.1.4

TID 4021

1.2.1.2.7.6

Area of Defined Region

1 cm2

TID 1401

1.2.1.2.7.6.1

Area Outline

SCOORD

TID 1401

1.2.1.2.7.6.1.1

Reference to node 1.1.4

TID 1401

Individual Impression/Recommendation 1.2.2 from Example 2 Content Tree

Figure E.3-9. Individual Impression/Recommendation 1.2.2 from Example 2 Content Tree


Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1.2.2

Individual Impression/Recommendation

TID 4003

1.2.2.1

Rendering Intent

Not for Presentation

TID 4003

1.2.2.2

Single Image Finding

Mammography breast density

TID 4006

1.2.2.2.1

Rendering Intent

Not for Presentation

TID 4006

1.2.2.2.2

Algorithm Name

"Density Detector"

TID 4019

1.2.2.2.3

Algorithm Version

"V3.7"

TID 4019

1.2.2.2.4

Center

POINT

TID 4021

1.2.2.2.4.1

Reference to node 1.1.2

TID 4021

1.2.2.2.5

Outline

SCOORD

TID 4021

1.2.2.2.5.1

Reference to node 1.1.2

TID 4021

Individual Impression/Recommendation 1.2.3 from Example 2 Content Tree

Figure E.3-10. Individual Impression/Recommendation 1.2.3 from Example 2 Content Tree


Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1.2.3

Individual Impression/Recommendation

TID 4003

1.2.3.1

Rendering Intent

Presentation Required

TID 4003

1.2.3.2

Single Image Finding

Calcification Cluster

TID 4006

1.2.3.2.1

Rendering Intent

Presentation Required

TID 4006

1.2.3.2.2

Algorithm Name

"Calc Cluster Detector"

TID 4019

1.2.3.2.3

Algorithm Version

"V2.4"

TID 4019

1.2.3.2.4

Center

POINT

TID 4021

1.2.3.2.4.1

Reference to node 1.1.3

TID 4021

1.2.3.2.5

Outline

SCOORD

TID 4021

1.2.3.2.5.1

Reference to node 1.1.3

TID 4021

1.2.3.2.6

Number of Calcifications

20

TID 4010

Individual Impression/Recommendation 1.2.4 from Example 2 Content Tree

Figure E.3-11. Individual Impression/Recommendation 1.2.4 from Example 2 Content Tree


Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1.2.4

Individual Impression/Recommendation

TID 4003

1.2.4.1

Rendering Intent

Presentation Required

TID 4003

1.2.4.2

Single Image Finding

Calcification Cluster

TID 4006

1.2.4.2.1

Rendering Intent

Presentation Required

TID 4006

1.2.4.2.2

Algorithm Name

"Calc Clustering"

TID 4019

1.2.4.2.3

Algorithm Version

"V2.4"

TID 4019

1.2.4.2.4

Center

POINT

TID 4021

1.2.4.2.4.1

Reference to node 1.1.1

TID 4021

1.2.4.2.5

Outline

SCOORD

TID 4021

1.2.4.2.5.1

Reference to node 1.1.1

TID 4021

1.2.4.2.6

Number of Calcifications

2

TID 4010

Single Image Finding 1.2.4.2.7 from Example 2 Content Tree

Figure E.3-12. Single Image Finding 1.2.4.2.7 from Example 2 Content Tree


Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1.2.4.2.7

Single Image Finding

Individual Calcification

TID 4006

1.2.4.2.7.1

Rendering Intent

Presentation Optional

TID 4006

1.2.4.2.7.2

Algorithm Name

"Calc Detector"

TID 4019

1.2.4.2.7.3

Algorithm Version

"V2.4"

TID 4019

1.2.4.2.7.4

Center

POINT

TID 4021

1.2.4.2.7.4.1

Reference to node 1.1.1

TID 4021

1.2.4.2.7.5

Outline

SCOORD

TID 4021

1.2.4.2.7.5.1

Reference to node 1.1.1

TID 4021

Single Image Finding 1.2.4.2.8 from Example 2 Content Tree

Figure E.3-13. Single Image Finding 1.2.4.2.8 from Example 2 Content Tree


Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1.2.4.2.8

Single Image Finding

Individual Calcification

TID 4006

1.2.4.2.8.1

Rendering Intent

Presentation Optional

TID 4006

1.2.4.2.8.2

Algorithm Name

"Calc Detector"

TID 4019

1.2.4.2.8.3

Algorithm Version

"V2.4"

TID 4019

1.2.4.2.8.4

Center

POINT

TID 4021

1.2.4.2.8.4.1

Reference to node 1.1.1

TID 4021

1.2.4.2.8.5

Outline

SCOORD

TID 4021

1.2.4.2.8.5.1

Reference to node 1.1.1

TID 4021

Summary of Detections Branch of Example 2 Content Tree

Figure E.3-14. Summary of Detections Branch of Example 2 Content Tree


Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1.3

Summary of Detections

Succeeded

TID 4000

1.3.1

Successful Detections

TID 4015

1.3.1.1

Detection Performed

Mammography breast density

TID 4017

1.3.1.1.1

Algorithm Name

"Density Detector"

TID 4019

1.3.1.1.2

Algorithm Version

"V3.7"

TID 4019

1.3.1.1.3

Reference to node 1.1.1

TID 4017

1.3.1.1.4

Reference to node 1.1.2

TID 4017

1.3.1.1.5

Reference to node 1.1.3

TID 4017

1.3.1.1.6

Reference to node 1.1.4

TID 4017

1.3.1.2

Detection Performed

Individual Calcification

TID 4017

1.3.1.2.1

Algorithm Name

"Calc Detector"

TID 4019

1.3.1.2.2

Algorithm Version

"V2.4"

TID 4019

1.3.1.2.3

Reference to node 1.1.1

TID 4017

1.3.1.2.4

Reference to node 1.1.2

TID 4017

1.3.1.2.5

Reference to node 1.1.3

TID 4017

1.3.1.2.6

Reference to node 1.1.4

TID 4017

1.3.1.3

Detection Performed

Calcification Cluster

TID 4017

1.3.1.3.1

Algorithm Name

"Calc Clustering"

TID 4019

1.3.1.3.2

Algorithm Version

"V2.4"

TID 4019

1.3.1.3.3

Reference to node 1.1.1

TID 4017

1.3.1.4

Detection Performed

Calcification Cluster

TID 4017

1.3.1.4.1

Algorithm Name

"Calc Cluster Detector"

TID 4019

1.3.1.4.2

Algorithm Version

"V2.4"

TID 4019

1.3.1.4.3

Reference to node 1.1.1

TID 4017

1.3.1.4.4

Reference to node 1.1.2

TID 4017

1.3.1.4.5

Reference to node 1.1.3

TID 4017

1.3.1.4.6

Reference to node 1.1.4

TID 4017

Summary of Analyses Branch of Example 2 Content Tree

Figure E.3-15. Summary of Analyses Branch of Example 2 Content Tree


Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1.4

Summary of Analyses

Succeeded

TID 4000

1.4.1

Successful Analyses

TID 4016

1.4.1.1

Analysis Performed

Mass Correlation

TID 4018

1.4.1.1.1

Algorithm Name

"Mass Maker"

TID 4019

1.4.1.1.2

Algorithm Version

"V1.9"

TID 4019

1.4.1.1.3

Reference to node 1.1.2

TID 4018

1.4.1.1.4

Reference to node 1.1.4

TID 4018

DICOM PS3.17 2019b - Explanatory Information