DICOM PS3.17 2019b - Explanatory Information

EE Relevant Patient Information Query (Informative)

EE.1 Relevant Patient Information Query Example (Informative)

The following is a simple and non-comprehensive example of a C-FIND Request for the Relevant Patient Information Query Service Class, specifically for the Breast Imaging Relevant Patient Information Query SOP Class, requesting a specific Patient ID, and requiring that any matching response be structured in the form of TID 9000 “Relevant Patient Information for Breast Imaging”.

C-FIND Request:

SR Tree Depth

Nesting

Attribute

Tag

VR

VL (hex)

Value

Affected SOP Class UID

(0000,0002)

UI

0018

1.2.840.10008.5.1.4.37.2

Command Field

(0000,0100)

US

0002

0020H [C-FIND-RQ]

Message ID

(0000,0110)

US

0002

0010H

Priority

(0000,0700)

US

0002

0000H [MEDIUM]

Data Set Type

(0000,0800)

US

0002

0102H

Patient's Name

(0010,0010)

PN

0000

Patient ID

(0010,0020)

LO

0008

MR975311

Patient's Birth Date

(0010,0030)

DA

0000

Patient's Sex

(0010,0040)

CS

0000

Observation DateTime

(0040,A032)

DT

0000

1

Value Type

(0040,A040)

CS

0000

1

Concept Name Code Sequence

(0040,A043)

SQ

0000

Content Template Sequence

(0040,A504)

SQ

ffffffff

%item

>

Mapping Resource

(0008,0105)

CS

0004

DCMR

>

Template Identifier

(0040,DB00)

CS

0004

9000

%enditem

%endseq

1

Content Sequence

(0040,A730)

SQ

0000

The following is a simple and non-comprehensive example of a C-FIND Response for the Relevant Patient Information Query Service Class, answering the C-FIND Request listed above, and structured in the form of TID 9000 “Relevant Patient Information for Breast Imaging” as required by the Affected SOP Class.

C-FIND Response #1:

SR Tree Depth

Nesting

Attribute

Tag

VR

VL (hex)

Value

Affected SOP Class UID

(0000,0002)

UI

0018

1.2.840.10008.5.1.4.37.2

Command Field

(0000,0100)

US

0002

8020H [C-FIND-RSP]

Message ID Being Responded To

(0000,0120)

US

0002

0010H

Data Set Type

(0000,0800)

US

0002

0102H

Status

(0000,0900)

US

0002

FF00H [Pending]

Patient's Name

(0010,0010)

PN

0008

Doe^Jane

Patient ID

(0010,0020)

LO

0008

MR975311

Patient's Birth Date

(0010,0030)

DA

0008

19541106

Patient's Sex

(0010,0040)

CS

0002

F

Observation DateTime

(0040,A032)

DT

000E

20021114124623

1

Value Type

(0040,A040)

CS

000A

CONTAINER

1

Concept Name Code Sequence

(0040,A043)

SQ

ffffffff

1

%item

1

>

Code Value

(0008,0100)

SH

0006

111511

1

>

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

1

>

Code Meaning

(0008,0104)

LO

0030

Relevant Patient Information for Breast Imaging

1

%enditem

1

%endseq

Content Template Sequence

(0040,A504)

SQ

ffffffff

%item

>

Mapping Resource

(0008,0105)

CS

0004

DCMR

>

Template Identifier

(0040,DB00)

CS

0004

9000

%enditem

%endseq

1

Content Sequence

(0040,A730)

SQ

ffffffff

1.1

%item

1.1

>

Relationship Type

(0040,A010)

CS

0010

HAS CONCEPT MOD

1.1

>

Value Type

(0040,A040)

CS

0004

CODE

1.1

>

Concept Name Code Sequence

(0040,A043)

SQ

ffffffff

1.1

%item

1.1

>>

Code Value

(0008,0100)

SH

0006

121049

1.1

>>

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

1.1

>>

Code Meaning

(0008,0104)

LO

0018

Language of Content Item and Descendants

1.1

%enditem

1.1

%endseq

1.1

>

Concept Code Sequence

(0040,A168)

SQ

ffffffff

1.1

%item

1.1

>>

Code Value

(0008,0100)

SH

0002

en

1.1

>>

Coding Scheme Designator

(0008,0102)

SH

0008

RFC3066

1.1

>>

Code Meaning

(0008,0104)

LO

0008

English

1.1

%enditem

1.1

%endseq

1.1

%enditem

1.2

%item

1.2

>

Relationship Type

(0040,A010)

CS

0008

CONTAINS

1.2

>

Value Type

(0040,A040)

CS

0004

NUM

1.2

>

Concept Name Code Sequence

(0040,A043)

SQ

ffffffff

1.2

%item

1.2

>>

Code Value

(0008,0100)

SH

0006

121033

1.2

>>

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

1.2

>>

Code Meaning

(0008,0104)

LO

000C

Subject Age

1.2

%enditem

1.2

%endseq

1.2

>

Measured Value Sequence

(0040,A300)

SQ

ffffffff

1.2

%item

1.2

>>

Measurement Units Code Sequence

(0040,08EA)

SQ

ffffffff

1.2

%item

1.2

>>>

Code Value

(0008,0100)

SH

0002

a

1.2

>>>

Coding Scheme Designator

(0008,0102)

SH

0004

UCUM

1.2

>>>

Coding Scheme Version

(0008,0103)

SH

0004

1.4

1.2

>>>

Code Meaning

(0008,0104)

LO

0004

Year

1.2

%enditem

1.2

%endseq

1.2

>>

Numeric Value

(0040,A30A)

DS

0002

48

1.2

%enditem

1.2

%endseq

1.2

%enditem

1.3

%item

1.3

>

Relationship Type

(0040,A010)

CS

0008

CONTAINS

1.3

>

Value Type

(0040,A040)

CS

000A

CONTAINER

1.3

>

Concept Name Code Sequence

(0040,A043)

SQ

ffffffff

1.3

%item

1.3

>>

Code Value

(0008,0100)

SH

000A

267011001

1.3

>>

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

1.3

>>

Code Meaning

(0008,0104)

LO

0016

Gynecological History

1.3

%enditem

1.3

%endseq

1.3

>

Continuity of Content

(0040,A050)

CS

0008

SEPARATE

1.3

>

Content Sequence

(0040,A730)

SQ

ffffffff

1.3.1

%item

1.3.1

>>

Relationship Type

(0040,A010)

CS

0008

CONTAINS

1.3.1

>>

Value Type

(0040,A040)

CS

0004

NUM

1.3.1

>>

Concept Name Code Sequence

(0040,A043)

SQ

ffffffff

1.3.1

%item

1.3.1

>>>

Code Value

(0008,0100)

SH

0006

111519

1.3.1

>>>

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

1.3.1

>>>

Code Meaning

(0008,0104)

LO

0020

Age at First Full Term Pregnancy

1.3.1

%enditem

1.3.1

%endseq

1.3.1

>>

Measured Value Sequence

(0040,A300)

SQ

ffffffff

1.3.1

%item

1.3.1

>>>

Measurement Units Code Sequence

(0040,08EA)

SQ

ffffffff

1.3.1

%item

1.3.1

>>>>

Code Value

(0008,0100)

SH

0002

a

1.3.1

>>>>

Coding Scheme Designator

(0008,0102)

SH

0004

UCUM

1.3.1

>>>>

Coding Scheme Version

(0008,0103)

SH

0004

1.4

1.3.1

>>>>

Code Meaning

(0008,0104)

LO

0004

Year

1.3.1

%enditem

1.3.1

%endseq

1.3.1

>>>

Numeric Value

(0040,A30A)

DS

0002

28

1.3.1

%enditem

1.3.1

%endseq

1.3.1

%enditem

1.3.2

%item

1.3.2

>>

Relationship Type

(0040,A010)

CS

0008

CONTAINS

1.3.2

>>

Value Type

(0040,A040)

CS

0004

NUM

1.3.2

>>

Concept Name Code Sequence

(0040,A043)

SQ

ffffffff

1.3.2

%item

1.3.2

>>>

Code Value

(0008,0100)

SH

0008

11977-6

1.3.2

>>>

Coding Scheme Designator

(0008,0102)

SH

0002

LN

1.3.2

>>>

Code Meaning

(0008,0104)

LO

0004

Para

1.3.2

%enditem

1.3.2

%endseq

1.3.2

>>

Measured Value Sequence

(0040,A300)

SQ

ffffffff

1.3.2

%item

1.3.2

>>>

Measurement Units Code Sequence

(0040,08EA)

SQ

ffffffff

1.3.2

%item

1.3.2

>>>>

Code Value

(0008,0100)

SH

0002

1

1.3.2

>>>>

Coding Scheme Designator

(0008,0102)

SH

0004

UCUM

1.3.2

>>>>

Coding Scheme Version

(0008,0103)

SH

0004

1.4

1.3.2

>>>>

Code Meaning

(0008,0104)

LO

0006

Unity

1.3.2

%enditem

1.3.2

%endseq

1.3.2

>>>

Numeric Value

(0040,A30A)

DS

0002

2

1.3.2

%enditem

1.3.2

%endseq

1.3.2

%enditem

1.3

%endseq

1.3

%enditem

1.4

%item

1.4

>

Relationship Type

(0040,A010)

CS

0008

CONTAINS

1.4

>

Value Type

(0040,A040)

CS

000A

CONTAINER

1.4

>

Concept Name Code Sequence

(0040,A043)

SQ

ffffffff

1.4

%item

1.4

>>

Code Value

(0008,0100)

SH

0006

111513

1.4

>>

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

1.4

>>

Code Meaning

(0008,0104)

LO

001C

Relevant Previous Procedures

1.4

%enditem

1.4

%endseq

1.4

>

Continuity of Content

(0040,A050)

CS

0008

SEPARATE

1.4

>

Content Sequence

(0040,A730)

SQ

ffffffff

1.4.1

%item

1.4.1

>>

Relationship Type

(0040,A010)

CS

0008

CONTAINS

1.4.1

>>

Value Type

(0040,A040)

CS

0004

CODE

1.4.1

>>

Concept Name Code Sequence

(0040,A043)

SQ

ffffffff

1.4.1

%item

1.4.1

>>>

Code Value

(0008,0100)

SH

0006

111531

1.4.1

>>>

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

1.4.1

>>>

Code Meaning

(0008,0104)

LO

0012

Previous Procedure

1.4.1

%enditem

1.4.1

%endseq

1.4.1

>>

Concept Code Sequence

(0040,A168)

SQ

ffffffff

1.4.1

%item

1.4.1

>>>

Code Value

(0008,0100)

SH

000A

287572003

1.4.1

>>>

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

1.4.1

>>>

Code Meaning

(0008,0104)

LO

0010

Cyst aspiration

1.4.1

%enditem

1.4.1

%endseq

1.4.1

>>

Content Sequence

(0040,A730)

SQ

ffffffff

1.4.1.1

%item

1.4.1.1

>>>

Relationship Type

(0040,A010)

CS

000E

HAS PROPERTIES

1.4.1.1

>>>

Value Type

(0040,A040)

CS

0004

CODE

1.4.1.1

>>>

Concept Name Code Sequence

(0040,A043)

SQ

ffffffff

1.4.1.1

%item

1.4.1.1

>>>>

Code Value

(0008,0100)

SH

000A

272741003

1.4.1.1

>>>>

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

1.4.1.1

>>>>

Code Meaning

(0008,0104)

LO

000A

Laterality

1.4.1.1

%enditem

1.4.1.1

%endseq

1.4.1.1

>>>

Concept Code Sequence

(0040,A168)

SQ

ffffffff

1.4.1.1

%item

1.4.1.1

>>>>

Code Value

(0008,0100)

SH

0008

80248007

1.4.1.1

>>>>

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

1.4.1.1

>>>>

Code Meaning

(0008,0104)

LO

000C

Left breast

1.4.1.1

%enditem

1.4.1.1

%endseq

1.4.1.1

%enditem

1.4.1.2

%item

1.4.1.2

>>>

Relationship Type

(0040,A010)

CS

000E

HAS PROPERTIES

1.4.1.2

>>>

Value Type

(0040,A040)

CS

0004

DATETIME

1.4.1.2

>>>

Concept Name Code Sequence

(0040,A043)

SQ

ffffffff

1.4.1.2

%item

1.4.1.2

>>>>

Code Value

(0008,0100)

SH

0006

122146

1.4.1.2

>>>>

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

1.4.1.2

>>>>

Code Meaning

(0008,0104)

LO

0012

Procedure DateTime

1.4.1.2

%enditem

1.4.1.2

%endseq

1.4.1.2

>>>

DateTime

(0040,A120)

DT

0008

19990825

1.4.1.2

%enditem

1.4.1

%endseq

1.4.1

%enditem

1.4

%endseq

1.4

%enditem

1.5

%item

1.5

>

Relationship Type

(0040,A010)

CS

0008

CONTAINS

1.5

>

Value Type

(0040,A040)

CS

000A

CONTAINER

1.5

>

Concept Name Code Sequence

(0040,A043)

SQ

ffffffff

1.5

%item

1.5

>>

Code Value

(0008,0100)

SH

0006

111515

1.5

>>

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

1.5

>>

Code Meaning

(0008,0104)

LO

0016

Relevant Risk Factors

1.5

%enditem

1.5

%endseq

1.5

>

Continuity of Content

(0040,A050)

CS

0008

SEPARATE

1.5

>

Content Sequence

(0040,A730)

SQ

ffffffff

1.5.1

%item

1.5.1

>>

Relationship Type

(0040,A010)

CS

0008

CONTAINS

1.5.1

>>

Value Type

(0040,A040)

CS

0004

CODE

1.5.1

>>

Concept Name Code Sequence

(0040,A043)

SQ

ffffffff

1.5.1

%item

1.5.1

>>>

Code Value

(0008,0100)

SH

0008

80943009

1.5.1

>>>

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

1.5.1

>>>

Code Meaning

(0008,0104)

LO

000C

Risk factor

1.5.1

%enditem

1.5.1

%endseq

1.5.1

>>

Concept Code Sequence

(0040,A168)

SQ

ffffffff

1.5.1

%item

1.5.1

>>>

Code Value

(0008,0100)

SH

0006

111559

1.5.1

>>>

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

1.5.1

>>>

Code Meaning

(0008,0104)

LO

0024

Weak family history of breast cancer

1.5.1

%enditem

1.5.1

%endseq

1.5.1

>>

Content Sequence

(0040,A730)

SQ

ffffffff

1.5.1.1

%item

1.5.1.1

>>>

Relationship Type

(0040,A010)

CS

000E

INFERRED FROM

1.5.1.1

>>>

Value Type

(0040,A040)

CS

0004

CODE

1.5.1.1

>>>

Concept Name Code Sequence

(0040,A043)

SQ

ffffffff

1.5.1.1

%item

1.5.1.1

>>>>

Code Value

(0008,0100)

SH

0006

111537

1.5.1.1

>>>>

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

1.5.1.1

>>>>

Code Meaning

(0008,0104)

LO

001E

Family Member with Risk Factor

1.5.1.1

%enditem

1.5.1.1

%endseq

1.5.1.1

>>>

Concept Code Sequence

(0040,A168)

SQ

ffffffff

1.5.1.1

%item

1.5.1.1

>>>>

Code Value

(0008,0100)

SH

0008

25211005

1.5.1.1

>>>>

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

1.5.1.1

>>>>

Code Meaning

(0008,0104)

LO

0004

Aunt

1.5.1.1

%enditem

1.5.1.1

%endseq

1.5.1.1

%enditem

1.5.1

%endseq

1.5.1

%enditem

1.5

%endseq

1.5

%enditem

1

%endseq

C-FIND Response #2:

SR Tree Depth

Nesting

Attribute

Tag

VR

VL (hex)

Value

Affected SOP Class UID

(0000,0002)

UI

0018

1.2.840.10008.5.1.4.37.2

Command Field

(0000,0100)

US

0002

8020H [C-FIND-RSP]

Message ID Being Responded To

(0000,0120)

US

0002

0010H

Data Set Type

(0000,0800)

US

0002

0101H

Status

(0000,0900)

US

0002

0000H [Success]

DICOM PS3.17 2019b - Explanatory Information