DICOM PS3.16 2019b - Content Mapping Resource

TID 5227 Cardiac Ultrasound Fetal Summary Section

Comments and observations of the procedure of immediate clinical interest.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 5227. Cardiac Ultrasound Fetal Summary Section

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

DT (125008, DCM, "Fetus Summary")

1

M

2

>

HAS OBS CONTEXT

INCLUDE

DTID 1008 “Subject Context, Fetus”

1

MC

IF this Template is invoked more than once to describe more than one fetus

3

>

CONTAINS

CODE

EV (121071, DCM, "Finding")

1-n

MC

IF row 4 does not exist

BCID 12249 “Cardiac Ultrasound Fetal Summary Codes”

4

>

CONTAINS

TEXT

EV (121071, DCM, "Finding")

1-n

MC

IF row 3 does not exist

5

>

CONTAINS

CODE

EV (387713003, SCT, "Surgical Procedure")

1-n

U

BCID 12247 “Pediatric, Fetal and Congenital Cardiac Surgical Interventions”


DICOM PS3.16 2019b - Content Mapping Resource