DICOM PS3.16 2020c - Content Mapping Resource

TID 5005 Fetal Biometry Section

The Fetal Biometry Section Template is a container for common biometric groups.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 5005. Fetal Biometry Section

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

DT (125002, DCM, "Fetal Biometry")

1

M

2

>

HAS OBS CONTEXT

INCLUDE

DTID 1008 “Subject Context, Fetus”

1

MC

IF this Template is invoked more than once to describe more than one fetus

3

>

CONTAINS

INCLUDE

DTID 5008 “Fetal Biometry Group”

1-n

M

$BiometryType = MemberOf {DCID 12005 “Fetal Biometry Measurements”

$TargetSite = DCID 12020 “Fetal Biometry Anatomic Sites”


Content Item Descriptions

Row 3

The group of measurements. Only one group per biometry type.

DICOM PS3.16 2020c - Content Mapping Resource