DICOM PS3.16 2019c - Content Mapping Resource

TID 4107 Chest CAD Geometry

All geometry Template invocations require specification of either the location of the center of the object, the outline, or both. Geometry is a property of single image findings (see TID 4104 “Chest CAD Single Image Finding”) and composite features (see TID 4103 “Chest CAD Composite Feature Body”).

Type:

Non-Extensible

Order:

Significant

Root:

No

Table TID 4107. Chest CAD Geometry

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

SCOORD

EV (111010, DCM, "Center")

1

MC

At least one of rows 1, 4 shall be present.

GRAPHIC TYPE = {POINT}

2

>

SELECTED FROM

IMAGE

1

MC

XOR row 3

3

>

R-SELECTED FROM

IMAGE

1

MC

XOR row 2

Shall reference an IMAGE Content Item in the (111028, DCM, "Image Library")

4

SCOORD

EV (111041, DCM, "Outline")

1

MC

At least one of rows 1, 4 shall be present.

5

>

SELECTED FROM

IMAGE

1

MC

XOR row 6

Shall reference the same Content Item as row 2

6

>

R-SELECTED FROM

IMAGE

1

MC

XOR row 5

Shall reference the same Content Item as row 3


DICOM PS3.16 2019c - Content Mapping Resource