DICOM PS3.16 2020a - Content Mapping Resource

TID 4023 CAD Operating Points

This Template describes CAD operating points. The description is deliberately left flexible and optional to allow implementation at differing levels of complexity.

Type:

Non-Extensible

Order:

Significant

Root:

No

Table TID 4023. CAD Operating Points

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

HAS PROPERTIES

NUM

EV (111072, DCM, "Maximum CAD Operating Point")

1

M

UNITS = DT ([arb'U], UCUM, "arbitrary unit")

Value is restricted to being an integer

2

HAS PROPERTIES

NUM

EV (111092, DCM, "Recommended CAD Operating Point")

1

U

UNITS = DT ({0:n}, UCUM, "range: 0:n"), where n is the value specified in row 1

Value is restricted to being an integer

3

HAS PROPERTIES

CONTAINER

EV (111093, DCM, "CAD Operating Point Table")

1

U

4

>

CONTAINS

CODE

EV (122698, DCM, "X-Concept")

1

M

DCID 6048 “CAD Operating Point Axis Label”

5

>

CONTAINS

CODE

EV (122699, DCM, "Y-Concept")

1

M

DCID 6048 “CAD Operating Point Axis Label”

6

>

CONTAINS

NUM

EV (111071, DCM, "CAD Operating Point")

1-n

M

Number of instances of this row shall equal value of row 1, plus 1.

UNITS = DT ({0:n}, UCUM, "range: 0:n"), where n is the value of Row 1. Value is restricted to being an integer that is unique within the invocation of this Template.

7

>>

HAS PROPERTIES

TEXT

EV (111081, DCM, "CAD Operating Point Description")

1

U

8

>>

HAS PROPERTIES

NUM

The value of Row 4

1

U

9

>>

HAS PROPERTIES

NUM

The value of Row 5

1

U


Content Item Descriptions

Maximum CAD Operating Point

The maximum possible value of CAD Operating Point for this type of Detection Performed. No CAD Operating Point value recorded in the CAD Processing and Findings Summary sub-tree of the report for this type of Detection Performed shall exceed this value. The report may or may not contain Rendering Intent = "Presentation Optional" detections that are assigned the maximum value.

Recommended CAD Operating Point

A number indicating which of the CAD operating points is recommended by the creator of a CAD SR instance as the first operating point to be used when rendering the CAD SR instance contents. Subsequent changes to the displayed operating point are implementation dependent.

DICOM PS3.16 2020a - Content Mapping Resource