DICOM PS3.16 2020a - Content Mapping Resource

TID 4003 Mammography CAD Individual Impression/Recommendation

This Template collects an individual impression the mammography CAD device had for a lesion, non-lesion object, or correlation of related objects. The details of the impression and recommendation are expressed in the Mammography CAD Impression/Recommendation Body (see TID 4002 “Mammography CAD Impression/Recommendation Body”). The data from which the details are inferred are expressed in the Composite Features (see TID 4004 “Mammography CAD Composite Feature”) and/or Single Image Findings (see TID 4006 “Mammography CAD Single Image Finding”) of which there may be several.

The sub-tree headed by this Template is illustrated in Figure E.1-3 “Example of Individual Impression/Recommendation Levels of Mammography CAD SR Content Tree” in PS3.17 .

Type:

Non-Extensible

Order:

Significant

Root:

No

Table TID 4003. Mammography CAD Individual Impression/Recommendation

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (111034, DCM, "Individual Impression/ Recommendation")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (111056, DCM, "Rendering Intent")

1

M

DCID 6034 “Intended Use of CAD Output”

3

>

CONTAINS

INCLUDE

DTID 4002 “Mammography CAD Impression/Recommendation Body”

1

U

4

>

CONTAINS

INCLUDE

DTID 4004 “Mammography CAD Composite Feature”

1-n

MC

At least one of rows 4, 5 shall be present.

5

>

CONTAINS

INCLUDE

DTID 4006 “Mammography CAD Single Image Finding”

1-n

MC

At least one of rows 4, 5 shall be present.


Content Item Descriptions

Rendering Intent

This Content Item constrains the SCP receiving the Mammography CAD SR IOD in its use of the contents of this Template and its Target Content Items. Mammography CAD devices may opt to use data marked "Not for Presentation" or "Presentation Optional" as input to subsequent mammography CAD processing steps. Refer to Section O.2 “Structured Reporting Storage SOP Class SCU and SCP Behavior” in PS3.4 .

DICOM PS3.16 2020a - Content Mapping Resource