DICOM PS3.16 2021a - Content Mapping Resource

TID 3990 Two Dimensional Measurement Graph

Generic Template representing arbitrary two-dimensional graphs.

Table TID 3990. Parameters

Parameter Name

Parameter Usage

$MeasurementGraph

Describes what the graph is about

$X-Concept

Concept of the X-Axis of the graph

$Y-Concept

Concept of the Y-Axis of the graph

$X-AxisUnit

Unit of the x-axis data elements

$Y-AxisUnit

Unit of the y-axis data elements


Type:

Extensible

Order:

Significant

Root:

No

Table TID 3990. Two Dimensional Measurement Graph

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

$MeasurementGraph

1

M

2

>

CONTAINS

CODE

EV (122698, DCM, "X-Concept")

1

M

$X-Concept

3

>

CONTAINS

CODE

EV (122699, DCM, "Y-Concept")

1

M

$Y-Concept

4

>

CONTAINS

CONTAINER

1-n

MC

IF Row 7, 8, or 9 not present

5

>>

CONTAINS

NUM

$X-Concept

1

M

UNITS = $X-AxisUnit

6

>>

CONTAINS

NUM

$Y-Concept

1

M

UNITS = $Y-AxisUnit

7

>

CONTAINS

IMAGE

$MeasurementGraph

1

U

8

>

CONTAINS

WAVEFORM

$MeasurementGraph

1

U

9

>

CONTAINS

COMPOSITE

$MeasurementGraph

1

U


Content Item Descriptions

Row 4

No Concept Name is used for this container.

Rows 5-6

The X-Concept values shall be monotonically increasing.

Row 7

Secondary Capture Image containing a bitmap representation of the graph

Row 8

Waveform containing a representation of the graph

Row 9

Composite Object containing a rendered representation of the graph

DICOM PS3.16 2021a - Content Mapping Resource