DICOM PS3.16 2020b - Content Mapping Resource

TID 3910 Flow Quantification

Contains the flow quantification measurement results

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3910. Flow Quantification

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINS

CONTAINER

EV (59776-5, LN, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (111004, DCM, "Analysis Performed")

1

M

EV (122604, DCM, "Flow Quantification")

3

>

HAS OBS CONTEXT

INCLUDE

DTID 3929 “Cardiovascular Analysis Observation Context”

1

U

4

>

HAS OBS CONTEXT

DATETIME

EV (398201009, SCT, "Start DateTime")

1

M

5

>

HAS OBS CONTEXT

DATETIME

EV (397898000, SCT, "Stop DateTime")

1

M

6

>

CONTAINS

INCLUDE

DTID 3990 “Two Dimensional Measurement Graph”

1

U

$MeasurementGraph = EV (122667, DCM, "Blood velocity vs. time of cardiac cycle")

$X-Concept = EV (122666, DCM, "Time relative to R-wave peak")

$Y-Concept = EV (252064005, SCT, "Arterial Velocity")

$X-AxisUnits = DT (ms, UCUM, "ms")

$Y-AxisUnits = DT (cm/s, UCUM, "cm/s")

7

>

CONTAINS

NUM

EV (122642, DCM, "Velocity Encoding Minimum Value")

1

U

UNITS = DT (cm/s, UCUM, "cm/s")

8

>

CONTAINS

NUM

EV (122643, DCM, "Velocity Encoding Maximum Value")

1

U

UNITS = DT (cm/s, UCUM, "cm/s")

9

>

CONTAINS

CONTAINER

EV (125007, DCM, "Measurement Group")

1-n

M

10

>>

HAS CONCEPT MOD

TEXT

EV (363698007, SCT, "Finding Site")

1

MC

XOR row 11

11

>>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

MC

XOR row 10

12

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122207, DCM, "Blood velocity, peak")

$Units = DT (cm/s, UCUM, "cm/s")

13

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122205, DCM, "Blood velocity, mean")

$Units = DT (cm/s, UCUM, "cm/s")

14

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (58190003, SCT, "Blood Flow")

$Derivation = EV (373098007, SCT, "Mean")

$Units = DT (ml/s, UCUM, "ml/s")

15

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-2

U

$Measurement = EV (58190003, SCT, "Blood Flow")

$ModType = EV (260674002, SCT, "Direction of flow")

$ModValue = DCID 12221 “Flow Direction”

$Units = DT (ml/s, UCUM, "ml/s")

16

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122645, DCM, "Net Forward Volume")

$Units = DT (ml, UCUM, "ml")

17

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122645, DCM, "Net Forward Volume")

$ModType = EV (121425, DCM, "Index")

$ModValue = DT (8277-6, LN, "BSA")

$Units = DT (ml/m2, UCUM, "ml/m2")

18

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (397415007, SCT, "Vessel Lumen Cross-Sectional Area")

$Derivation = DCID 3488 “Min/Max/Mean”

$Units = DT (mm2, UCUM, "mm2")


DICOM PS3.16 2020b - Content Mapping Resource