DICOM PS3.16 2019a - Content Mapping Resource

TID 3902 Vascular Analysis

Contains either morphological or functional vascular measurement results of an analysis

Table TID 3902. Parameters

Parameter Name

Parameter Usage

$AnalysisPerformed

Analysis Performed


Type:

Extensible

Order:

Significant

Root:

No

Table TID 3902. Vascular Analysis

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (121070, DCM, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (111004, DCM, "Analysis Performed")

1

M

$AnalysisPerformed

3

>

CONTAINS

INCLUDE

DTID 3905 “Calcium Scoring Results”

1

UC

IFF the value of row 2 equals EV (122605, DCM, "Vascular Morphological Analysis")

4

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-D0767, SRT, "Blood Vessel of Head")

$SectionLaterality = EV (G-A101, SRT, "Left")

$Anatomy = DCID 12105 “Intracranial Cerebral Vessels”

$AnalysisPerformed = $AnalysisPerformed

5

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-D0767, SRT, "Blood Vessel of Head")

$SectionLaterality = EV (G-A100, SRT, "Right")

$Anatomy = DCID 12105 “Intracranial Cerebral Vessels”

$AnalysisPerformed = $AnalysisPerformed

6

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-D0767, SRT, "Blood Vessel of Head")

$SectionLaterality = EV (G-A103, SRT, "Unilateral")

$Anatomy = DCID 12106 “Intracranial Cerebral Vessels (Unilateral)”

$AnalysisPerformed = $AnalysisPerformed

7

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-45005, SRT, "Artery of Neck")

$SectionLaterality = EV (G-A101, SRT, "Left")

$Anatomy = DCID 12104 “Extracranial Arteries”

$AnalysisPerformed = $AnalysisPerformed

8

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-45005, SRT, "Artery of Neck")

$SectionLaterality = EV (G-A100, SRT, "Right")

$Anatomy = DCID 12104 “Extracranial Arteries”

$AnalysisPerformed = $AnalysisPerformed

9

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-47040, SRT, "Artery of Lower Extremity ")

$SectionLaterality = EV (G-A101, SRT, "Left")

$Anatomy = DCID 12109 “Lower Extremity Arteries”

$AnalysisPerformed = $AnalysisPerformed

10

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-47040, SRT, "Artery of Lower Extremity ")

$SectionLaterality = EV (G-A100, SRT, "Right")

$Anatomy = DCID 12109 “Lower Extremity Arteries”

$AnalysisPerformed = $AnalysisPerformed

11

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-49403, SRT, "Vein of Lower Extremity")

$SectionLaterality = EV (G-A101, SRT, "Left")

$Anatomy = DCID 12110 “Lower Extremity Veins”

$AnalysisPerformed = $AnalysisPerformed

12

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-49403, SRT, "Vein of Lower Extremity")

$SectionLaterality = EV (G-A100, SRT, "Right")

$Anatomy = DCID 12110 “Lower Extremity Veins”

$AnalysisPerformed = $AnalysisPerformed

13

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-47020, SRT, "Artery of Upper Extremity")

$SectionLaterality = EV (G-A101, SRT, "Left")

$Anatomy = DCID 12107 “Upper Extremity Arteries”

$AnalysisPerformed = $AnalysisPerformed

14

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-47020, SRT, "Artery of Upper Extremity")

$SectionLaterality = EV (G-A100, SRT, "Right")

$Anatomy = DCID 12107 “Upper Extremity Arteries”

$AnalysisPerformed = $AnalysisPerformed

15

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-49103, SRT, "Vein of Upper Extremity")

$SectionLaterality = EV (G-A101, SRT, "Left")

$Anatomy = DCID 12108 “Upper Extremity Veins”

$AnalysisPerformed = $AnalysisPerformed

16

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-49103, SRT, "Vein of Upper Extremity")

$SectionLaterality = EV (G-A100, SRT, "Right")

$Anatomy = DCID 12108 “Upper Extremity Veins”

$AnalysisPerformed = $AnalysisPerformed

17

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-71019, SRT, "Vascular Structure of Kidney")

$SectionLaterality = EV (G-A101, SRT, "Left")

$Anatomy = DCID 12115 “Renal Vessels”

$AnalysisPerformed = $AnalysisPerformed

18

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-71019, SRT, "Vascular Structure of Kidney")

$SectionLaterality = EV (G-A100, SRT, "Right")

$Anatomy = DCID 12115 “Renal Vessels”

$AnalysisPerformed = $AnalysisPerformed

19

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-46002, SRT, "Artery of Abdomen")

$SectionLaterality = EV (G-A101, SRT, "Left")

$Anatomy = DCID 12111 “Abdominal Arteries (Lateral)”

$AnalysisPerformed = $AnalysisPerformed

20

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-46002, SRT, "Artery of Abdomen")

$SectionLaterality = EV (G-A100, SRT, "Right")

$Anatomy = DCID 12111 “Abdominal Arteries (Lateral)”

$AnalysisPerformed = $AnalysisPerformed

21

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-46002, SRT, "Artery of Abdomen")

$SectionLaterality = EV (G-A103, SRT, "Unilateral")

$Anatomy = DCID 12112 “Abdominal Arteries (Unilateral)”

$AnalysisPerformed = $AnalysisPerformed

22

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-487A0, SRT, "Vein of Abdomen")

$SectionLaterality = EV (G-A101, SRT, "Left")

$Anatomy = DCID 12113 “Abdominal Veins (Lateral)”

$AnalysisPerformed = $AnalysisPerformed

23

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-487A0, SRT, "Vein of Abdomen")

$SectionLaterality = EV (G-A100, SRT, "Right")

$Anatomy = DCID 12113 “Abdominal Veins (Lateral)”

$AnalysisPerformed = $AnalysisPerformed

24

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-487A0, SRT, "Vein of Abdomen")

$SectionLaterality = EV (G-A103, SRT, "Unilateral")

$Anatomy = DCID 12114 “Abdominal Veins (Unilateral)”

$AnalysisPerformed = $AnalysisPerformed

25

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-44000, SRT, "Pulmonary Artery Structure")

$SectionLaterality = EV (G-A103, SRT, "Unilateral")

$Anatomy = DCID 3829 “Pulmonary Arteries”

$AnalysisPerformed = $AnalysisPerformed

26

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-43000, SRT, "Coronary Artery Structure")

$Anatomy = DCID 3015 “Coronary Arteries”

$AnalysisPerformed = $AnalysisPerformed

27

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-48400, SRT, "Cardiac Vein Structure")

$Anatomy = DCID 3839 “Coronary Veins”

$AnalysisPerformed = $AnalysisPerformed

28

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (T-48581, SRT, "Pulmonary Venous Structure")

$Anatomy = DCID 3840 “Pulmonary Veins”

$AnalysisPerformed = $AnalysisPerformed


DICOM PS3.16 2019a - Content Mapping Resource