DICOM PS3.16 2020c - Content Mapping Resource

TID 3717 ECG Qualitative Analysis

The ECG Qualitative Analysis Template allows a free text qualitative interpretation of the analyzed ECG, as well as a structure for a coded analysis.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3717. Qualitative Analysis, ECG

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINS

CONTAINER

EV (122145, DCM, "Qualitative Analysis")

1

M

2

>

CONTAINS

TEXT

EV (271921002, SCT, "ECG Finding")

1

MC

At least one of rows 2 and 3 shall be present

3

>

CONTAINS

CODE

EV (271921002, SCT, "ECG Finding")

1-n

MC

At least one of rows 2 and 3 shall be present

BCID 3230 “ECG Findings”

4

>>

HAS CONCEPT MOD

TEXT

EV (121051, DCM, "Equivalent Meaning of Value")

1

U

5

>>

INFERRED FROM

CODE

EV (271921002, SCT, "ECG Finding")

1-n

U

6

>>>

HAS CONCEPT MOD

TEXT

EV (121051, DCM, "Equivalent Meaning of Value")

1

U


Content Item Descriptions

Row 3-4

ECG Finding provides one or more coded interpretive statements using standard or implementation-specific codes. Each coded finding will include a Code Meaning (0008,0104) using the LO Value Representation (64 characters); longer human-readable text strings for interpretive statements may be conveyed in the Row 4 Equivalent Meaning of Value Content Item..

Row 5-6

Each primary ECG Finding of Row 3 may have multiple supporting coded findings in Row 5, with longer human-readable text strings for interpretive statements if necessary in Row 6.

DICOM PS3.16 2020c - Content Mapping Resource