DICOM PS3.16 2020c - Content Mapping Resource

TID 3713 ECG Global Measurements

The ECG Global Measurements Template provides a structure for measurements calculated across the ECG waveform as a whole (multiple leads).

As an Extensible Template, applications may include any ECG global measurements, such as angles of the electrical vector of various ECG waves. The recommended vocabulary for such concepts is ISO/IEEE 11073-10102.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3713. ECG Global Measurements

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (122158, DCM, "ECG Global Measurements")

1

M

2

>

HAS OBS CONTEXT

INCLUDE

DTID 3715 “ECG Measurement Source”

1

U

3

>

CONTAINS

NUM

DT (2:16020, MDC, "Atrial Heart Rate")

1

U

UNITS = EV ({H.B.}/min, UCUM, "BPM")

4

>

CONTAINS

NUM

DT (2:16016, MDC, "Ventricular Heart Rate")

1

M

UNITS = EV ({H.B.}/min, UCUM, "BPM")

5

>

CONTAINS

NUM

DT (2:16160, MDC, "QT interval global")

1

M

UNITS = EV (ms, UCUM, "ms")

6

>

CONTAINS

NUM

DT (2:15876, MDC, "QTc interval global")

1

U

UNITS = EV (ms, UCUM, "ms")

7

>>

HAS PROPERTIES

CODE

DT (111001, DCM, "Algorithm Name")

1

U

DCID 3678 “QT Correction Algorithms”

8

>

CONTAINS

NUM

DT (2:15872, MDC, "PR interval global")

1

M

UNITS = EV (ms, UCUM, "ms")

9

>

CONTAINS

NUM

DT (2:16156, MDC, "QRS duration global")

1

M

UNITS = EV (ms, UCUM, "ms")

10

>

CONTAINS

NUM

DT (2:16168, MDC, "RR interval global")

1

M

UNITS = EV (ms, UCUM, "ms")

11

>

CONTAINS

NUM

DCID 3689 “ECG Global Waveform Durations”

1-n

U

UNITS = EV (ms, UCUM, "ms")

12

>

CONTAINS

NUM

DCID 3229 “ECG Axis Measurements”

1-n

U

UNITS = EV (deg, UCUM, "deg")

13

>

CONTAINS

NUM

DT (2:16032, MDC, "Count of all beats")

1

U

UNITS = EV ({beats}, UCUM, "beats")

14

>

CONTAINS

NUM

DT (122707, DCM, "Number of Ectopic Beats")

1

U

UNITS = EV ({beats}, UCUM, "beats")

15

>>

HAS PROPERTIES

CODE

EV (116676008, SCT, "Associated Morphology")

1-n

U

BCID 3234 “Ectopic Beat Morphology”


DICOM PS3.16 2020c - Content Mapping Resource