DICOM PS3.16 2019b - Content Mapping Resource

TID 3501 Hemodynamics Measurement Group

The Hemodynamic Measurement Group Template provides a structure for measurements acquired during a single procedure phase in a cardiac catheterization lab.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3501. Hemodynamics Measurement Group

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (121070, DCM, "Findings")

1

M

2

>

HAS ACQ CONTEXT

CODE

EV (129085009, SCT, "Catheterization Procedure Phase")

1

M

DCID 3651 “Hemodynamic Measurement Phase”

3

>

HAS ACQ CONTEXT

INCLUDE

DTID 3520 “Hemodynamic Clinical Context”

1

U

4

>

HAS ACQ CONTEXT

TEXT

EV (121124, DCM, "Procedure Action ID")

1

U

5

>

CONTAINS

INCLUDE

DTID 3510 “Vital Signs”

1-n

U

6

>

CONTAINS

INCLUDE

DTID 3504 “Arterial Pressure Measurement”

1-n

U

7

>

CONTAINS

INCLUDE

DTID 3505 “Atrial Pressure Measurement”

1-n

U

8

>

CONTAINS

INCLUDE

DTID 3506 “Venous Pressure Measurement”

1-n

U

9

>

CONTAINS

INCLUDE

DTID 3507 “Ventricular Pressure Measurement”

1-n

U

10

>

CONTAINS

INCLUDE

DTID 3508 “Gradient Measurement”

1-n

U

11

>

CONTAINS

INCLUDE

DTID 3509 “Blood Velocity Measurement”

1-n

U

12

>

CONTAINS

INCLUDE

DTID 3515 “Cardiac Output Measurement by Indicator Dilution”

1-n

U

13

>

CONTAINS

INCLUDE

DTID 3516 “Blood Lab Measurements”

1-n

U

14

>

CONTAINS

INCLUDE

DTID 3560 “Derived Hemodynamic Measurements”

1-n

U

15

>

CONTAINS

INCLUDE

DTID 3714 “ECG Lead Measurements”

1-n

U


Content Item Descriptions

Row 4

Procedure Action ID allows linkage between the hemodynamic measurements recorded in this Template and a procedure step or phase recorded in the Procedure Log, e.g., using TID 3100 “Procedure Action”.