DICOM PS3.16 2019a - Content Mapping Resource

TID 3312 Physiological Summary

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3312. Physiological Summary

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUM

DT (40443-4, LN, "Resting Heart Rate")

1

M

UNITS = DT ({H.B.}/min, UCUM, "BPM")

2

NUM

EV (F-008EC, SRT, "Systolic Blood Pressure")

1

M

UNITS = DCID 3500 “Pressure Units”

3

>

HAS CONCEPT MOD

CODE

EV (109054, DCM, "Patient State")

1

M

DT (F-01604, SRT, "Resting State")

4

NUM

EV (F-008ED, SRT, "Diastolic Blood Pressure")

1

M

UNITS = DCID 3500 “Pressure Units”

5

>

HAS CONCEPT MOD

CODE

EV (109054, DCM, "Patient State")

1

M

DT (F-01604, SRT, "Resting State")

6

NUM

DT (F-04F92, SRT, "Target HR")

1

M

UNITS = DT ({H.B.}/min, UCUM, "BPM")

7

NUM

DT (F-04FA6, SRT, "Maximum HR Achieved")

1

M

UNITS = DT ({H.B.}/min, UCUM, "BPM")

8

NUM

DT (F-04FA6, SRT, "Maximum HR Achieved")

1

M

UNITS = EV (%, UCUM, "%")

9

>

HAS CONCEPT MOD

CODE

EV (121425, DCM, "Index")

1

M

DT (F-04F92, SRT, "Target HR")

10

NUM

DT (122716, DCM, "Maximum Power Output Achieved")

1

U

UNITS = DT (W, UCUM, "Watts")

11

NUM

DT (122717, DCM, "Peak activity workload")

1

U

UNITS = DT ([MET], UCUM, "METS")

12

CODE

DT (F-04F9F, SRT, "HR Response")

1

U

DCID 3210 “Speed of Response”

13

NUM

DT (F-00E11, SRT, "Maximum systolic blood pressure")

1

U

UNITS = DCID 3500 “Pressure Units”

14

NUM

DT (F-00E21, SRT, "Maximum diastolic blood pressure")

1

U

UNITS = DCID 3500 “Pressure Units”

15

CODE

DT (F-04F74, SRT, "BP Response")

1

U

DCID 3210 “Speed of Response”

16

NUM

DT (122718, DCM, "Peak Double Product")

1

U

UNITS = DT (mm[Hg].{H.B.}/min, UCUM, "mmHg.BPM")

17

NUM

DT (F-031F8, SRT, "Total Exercise duration")

1

U

UNITS = DT (min, UCUM, "min")

18

NUM

DT (F-031F7, SRT, "Total test duration")

1

U

UNITS = DT (min, UCUM, "min")

19

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DT (F-04FCC, SRT, "Functional capacity")

20

TEXT

DT (F-04FCC, SRT, "Functional capacity")

1

U

21

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (122760, DCM, "Stress test score")

$Method = BCID 3238 “Stress Scoring Scales”

22

NUM

DT (F-04FCA, SRT, "Heart rate recovery time")

1

U

UNITS = DT (s, UCUM, "s")

23

CODE

EV (121071, DCM, "Finding")

1-n

U

BCID 3213 “Stress Hemodynamic Findings”

24

CODE

EV (F-00F4E, SRT, "Cardiovascular event risk")

1

U

BCID 3118 “Level of Risk”


Content Item Descriptions

Row 22

Numerical scoring of a patient's functional capacity shall include the range of the scoring system in the Units of Measurement (see Section 7.2.2), and may include a coded identifier for the scoring system in the Method concept modifier of TID 300 “Measurement”.

DICOM PS3.16 2019a - Content Mapping Resource