DICOM PS3.16 2019a - Content Mapping Resource

TID 3304 Stress Test Measurement Group

Each instance of the Stress Test Measurement Group represents a group of data elements acquired at approximately the same instant, and conventionally rendered as row in a tabular presentation. It is typically generated during the Stress exam whenever a time interval elapses (for example, every minute of the phase), when a technician observes data worth capturing, or when measurements exceed a given range.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3304. Stress Test Measurement Group

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

DT (121070, DCM, "Findings")

1

M

2

>

CONTAINS

NUM

DT (F-031F9, SRT, "Time since start of exam")

1

M

UNITS = DT (min, UCUM, "min")

3

>

CONTAINS

NUM

DT (122710, DCM, "Time since start of stage")

1

M

UNITS = DT (min, UCUM, "min")

4

>

CONTAINS

NUM

DT (122702, DCM, "Treadmill speed")

1

U

UNITS = DCID 3212 “Treadmill Speed”

5

>

CONTAINS

NUM

DT (122703, DCM, "Treadmill gradient")

1

U

UNITS = EV (%, UCUM, "%")

6

>

CONTAINS

NUM

DT (122704, DCM, "Ergometer power")

1

U

UNITS = EV (W, UCUM, "Watts")

7

>

CONTAINS

NUM

DT (122709, DCM, "Activity workload")

1

U

UNITS = DT ([MET], UCUM, "METS")

8

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DT (122706, DCM, "Rating of Perceived Exertion")

$Method = BCID 3239 “Perceived Exertion Scales”

9

>

CONTAINS

NUM

DT (122705, DCM, "Pharmacological Stress Agent Dose Rate")

1

MC

IFF Pharmacological Stress used

UNITS = EV (ug/kg/min, UCUM, "ug/kg/min")

10

>

CONTAINS

INCLUDE

DTID 3106 “Drugs/Contrast Administered”

1

U

11

>

CONTAINS

NUM

EV (8867-4, LN, "Heart Rate")

1

U

UNITS = EV ({H.B.}/min, UCUM, "BPM")

12

>

CONTAINS

NUM

EV (F-008EC, SRT, "Systolic Blood Pressure")

1

U

UNITS = DCID 3500 “Pressure Units”

13

>

CONTAINS

NUM

EV (F-008ED, SRT, "Diastolic Blood Pressure")

1

U

UNITS = DCID 3500 “Pressure Units”

14

>

CONTAINS

NUM

DT (122707, DCM, "Number of Ectopic Beats")

1

U

UNITS = EV ({beats}, UCUM, "beats")

15

>>

HAS PROPERTIES

NUM

DT (R-40861, SRT, "Period of collection")

1

M

UNITS = DT (min, UCUM, "min")

16

>>

HAS PROPERTIES

CODE

EV (G-C504, SRT, "Associated Morphology")

1-n

U

BCID 3234 “Ectopic Beat Morphology”

17

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DT (F-03204, SRT, "ST Elevation")

$Units = DT (mV, UCUM, "mV")

$TargetSite = DCID 3001 “ECG Leads”

18

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DT (F-38279, SRT, "ST Depression")

$Units = DT (mV, UCUM, "mV")

$TargetSite = DCID 3001 “ECG Leads”

19

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3228 “ECG Timing Measurements”

$Units = DT (ms, UCUM, "ms")

$TargetSite = DCID 3001 “ECG Leads”

20

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3227 “QTc Measurements”

$Units = DT (ms, UCUM, "ms")

$TargetSite = DCID 3001 “ECG Leads”

$Equation = DCID 3678 “QT Correction Algorithms”

21

>>

INFERRED FROM

NUM

DT (2:16000, MDC, "RR Interval for QTc")

1

U

UNITS = DT (ms, UCUM, "ms")

22

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3229 “ECG Axis Measurements”

$Units = DT (deg, UCUM, "deg")

23

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3526 “Blood Gas Saturation”

$Units = EV (%, UCUM, "%")

24

>

CONTAINS

NUM

DT (122708, DCM, "Double Product")

1

U

UNITS = DT (mm[Hg].{H.B.}/min, UCUM, "mmHg.BPM")

25

>

CONTAINS

CODE

EV (121071, DCM, "Finding")

1-n

U

DCID 3220 “Stress Symptoms”

26

>

CONTAINS

CODE

EV (F-00033, SRT, "ECG Finding")

1-n

U

DCID 3230 “ECG Findings”

27

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1

U


Content Item Descriptions

Row 1

The Container shall have a specific Content Item Observation DateTime (0040,A032) attribute to indicate the time at which the measurements were made.

Row 10

Included TID 3106 “Drugs/Contrast Administered” allows the recording of test medications other than the Pharmacological Stress Agent identified in TID 3301 “Stress Test Procedure Description”.

Rows 17, 18

ECG ST elevation/depression is measured in units of mV, but is conventionally reported in units of mm, based on strip recordings with scaling of 100 uV/mm. The display application should render these measurements in units meaningful to the user.

Row 19

Note that the MDC codes for "per lead" measurements specified in CID 3228 “ECG Timing Measurements” are base codes for post-coordination with lead identifiers conveyed in the Target Site modifier in TID 300 “Measurement”. MDC also defines pre-coordinated codes that include both the measurement and the lead, which may be used in this row.

Row 20

Note that the MDC code for "QTc interval per lead" specified in CID 3227 “QTc Measurements” is a base code for post-coordination with lead identifiers conveyed in the Target Site modifier in TID 300 “Measurement”. MDC also defines pre-coordinated codes that include both the measurement and the lead, which may be used in this row.

Note that TID 300 “Measurement” enables the encoding of a non-standard correction algorithm, either as a local code, or as a TEXT Method Citation (see TID 300 “Measurement” row 12).

Row 21

R-R interval used for QT correction algorithm

Row 22

Recommended range for ECG axis measurements is -90° to +270°

DICOM PS3.16 2019a - Content Mapping Resource