DICOM PS3.16 2019a - Content Mapping Resource

TID 3218 Position in Arterial Segment

The Position in Arterial Segment Template consists of the position Content Items common for the Angiographic Lesion Analysis and Sub-Segmental Data.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3218. Position in Arterial Segment

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122528, DCM, "Position of Proximal Border")

$Unit = DT (mm, UCUM, "mm")

2

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122529, DCM, "Position of Distal Border")

$Unit = DT (mm, UCUM, "mm")

3

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122382, DCM, "Site of Luminal Minimum")

$Unit = DT (mm, UCUM, "mm")

4

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122516, DCM, "Site of Luminal Maximum")

$Unit = DT (mm, UCUM, "mm")

5

CONTAINS

NUM

EV (122528, DCM, "Position of Proximal Border")

1

UC

IFF TID 3214 “Analyzed Segment” Row 14 is present

UNITS = DT ({pixels}, UCUM, "pixels")

6

CONTAINS

NUM

EV (122529, DCM, "Position of Distal Border")

1

UC

IFF TID 3214 “Analyzed Segment” Row 14 is present

UNITS = DT ({pixels}, UCUM, "pixels")

7

CONTAINS

NUM

EV (122382, DCM, "Site of Luminal Minimum")

1

UC

IFF TID 3214 “Analyzed Segment” Row 14 is present

UNITS = DT ({pixels}, UCUM, "pixels")

8

CONTAINS

NUM

EV (122516, DCM, "Site of Luminal Maximum")

1

UC

IFF TID 3214 “Analyzed Segment” Row 14 is present

UNITS = DT ({pixels}, UCUM, "pixels")


Content Item Descriptions

Row 1, 2, 3, 4

Positions are relative to the midpoint of the first left and right contour points and measured along the midline of the left and right contour

Row 5, 6, 7, 8

The positions are relative to the measurement locations of the Diameter Graph of TID 3214 “Analyzed Segment” row 14.

DICOM PS3.16 2019a - Content Mapping Resource