DICOM PS3.16 2020b - Content Mapping Resource

TID 3208 Frame-to-Frame Results

The Frame-to-Frame Result Template consists of a CONTAINER providing measurements derived from the angiographic images on frame-by-frame basis.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3208. Frame-To-Frame Result

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (59776-5, LN, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (111004, DCM, "Analysis Performed")

1

M

EV (122499, DCM, "Frame to Frame Analysis")

3

>

CONTAINS

IMAGE

EV (121112, DCM, "Source of Measurement")

1-2

M

VM = 1: Single plane analysis, VM = 2: Biplane analysis

4

>

CONTAINS

CODE

EV (122429, DCM, "Volume Method")

1

M

DCID 3453 “Cardiac Volume Methods”

5

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

M

$Measurement = DCID 3471 “Estimated Volumes”

$TargetSite = DCID 3462 “Chamber Identification”

$Unit = DT (ml, UCUM, "ml")

6

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (122445, DCM, "Wall Thickness")

$Unit = DT (mm, UCUM, "mm")

7

>

CONTAINS

IMAGE

1-n

U


Content Item Descriptions

Row 3

Identifies each frame analyzed, using the multi-valued Referenced Frame Number (0008,1160) attribute of the IMAGE Content Item.

Row 5, 6

Includes one measurement for each frame referenced in Row 3.

Row 7

Secondary Capture image with ventricular contours. No purpose of reference is specified.

DICOM PS3.16 2020b - Content Mapping Resource