DICOM PS3.16 2019a - Content Mapping Resource

TID 3207 AA Main Results

The AA Main Results Template consists of a CONTAINER with a structure for reporting the main atrial analysis measurements.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3207. AA Main Results

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (121070, DCM, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (G-C0E3, SRT, "Finding Site")

1

M

DCID 3462 “Chamber Identification”

3

>

CONTAINS

CODE

EV (122429, DCM, "Volume Method")

1

M

DCID 3453 “Cardiac Volume Methods”

4

>

CONTAINS

NUM

EV (122435, DCM, "Regression Volume Exponent")

1

U

Unit = DT (1, UCUM, "no units")

5

>

CONTAINS

NUM

EV (122431, DCM, "Regression Slope ED")

1

U

Unit = DT ({ratio}, UCUM, "ratio")

6

>

CONTAINS

NUM

EV (122432, DCM, "Regression Offset ED")

1

U

Unit = DT (ml, UCUM, "ml")

7

>

CONTAINS

NUM

EV (122433, DCM, "Regression Slope ES")

1

U

Unit = DT ({ratio}, UCUM, "ratio")

8

>

CONTAINS

NUM

EV (122434, DCM, "Regression Offset ES")

1

U

Unit = DT (ml, UCUM, "ml")

9

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3468 “ED Volume”

$Unit = DT (ml, UCUM, "ml")

10

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3469 “ES Volume”

$Unit = DT (ml, UCUM, "ml")

11

>

CONTAINS

IMAGE

No purpose of reference

1-n

U


Content Item Descriptions

Row 11

Secondary Capture image with ED and/or ES contours

DICOM PS3.16 2019a - Content Mapping Resource