DICOM PS3.16 2020c - Content Mapping Resource

TID 3112 Specimen Obtained

The Specimen Obtained Template allows recording of obtaining a specimen, and the identifiers for that specimen. This is particularly designed for blood samples that will be analyzed for blood oxygen-related measurements. The analysis of the sample may be recorded in one or more log entries using TID 3109 “Measurements”, or in a separate Structured Report SOP Instance referenced by a log entry using TID 3103 “Referenced Object”.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3112. Specimen Obtained

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (121123, DCM, "Patient Status or Event")

1

M

DCID 3515 “Specimen Collection”

2

>

HAS ACQ CONTEXT

CODE

EV (371439000, SCT, "Specimen Type")

1

UC

IFF specimen is blood sample

DCID 3520 “Blood Source Type”

3

>

HAS ACQ CONTEXT

CODE

EV (363704007, SCT, "Procedure site")

1

U

BCID 3630 “Cardiovascular Anatomic Locations”

4

>

HAS PROPERTIES

INCLUDE

DTID 1009 “Subject Context, Specimen”

1

U


DICOM PS3.16 2020c - Content Mapping Resource