DICOM PS3.16 2019a - Content Mapping Resource

TID 10003C Irradiation Event X-Ray Mechanical Data

This Template contains data that is expected to be available to the gantry or mechanical component of the equipment.

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 10003C. Irradiation Event X-Ray Mechanical Data

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (113956, DCM, "CR/DR Mechanical Configuration")

1

U

DCID 10031 “CR/DR Mechanical Configuration”

2

NUM

EV (112011, DCM, "Positioner Primary Angle")

1

UC

XOR Row 6

UNITS = EV (deg, UCUM, "deg")

3

NUM

EV (112012, DCM, "Positioner Secondary Angle")

1

UC

XOR Row 6

UNITS = EV (deg, UCUM, "deg")

4

NUM

EV (113739, DCM, "Positioner Primary End Angle")

1

UC

IFF TID 10003 Row 7 value = (113613, DCM, "Rotational Acquisition")

UNITS = EV (deg, UCUM, "deg")

5

NUM

EV (113740, DCM, "Positioner Secondary End Angle")

1

UC

IFF TID 10003 Row 7 value = (113613, DCM, "Rotational Acquisition")

UNITS = EV (deg, UCUM, "deg")

6

NUM

EV (113770, DCM, "Column Angulation")

1

UC

XOR Rows 2, 3

UNITS = EV (deg, UCUM, "deg")

7

NUM

EV (113754, DCM, "Table Head Tilt Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

8

NUM

EV (113755, DCM, "Table Horizontal Rotation Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

9

NUM

EV (113756, DCM, "Table Cradle Tilt Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

10

NUM

EV (111633, DCM, "Compression Thickness")

1

U

UNITS = EV (mm, UCUM, "mm")

11

NUM

DCID 10008 “Dose Related Distance Measurements”

1-n

U

UNITS = EV (mm, UCUM, "mm")

12

NUM

EV (128757, DCM, "Positioner Isocenter Primary Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

13

NUM

EV (128758, DCM, "Positioner Isocenter Secondary Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

14

NUM

EV (128759, DCM, "Positioner Isocenter Detector Rotation Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

15

NUM

EV (128760, DCM, "Positioner Isocenter Primary End Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

16

NUM

EV (128761, DCM, "Positioner Isocenter Secondary End Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

17

NUM

EV (128762, DCM, "Positioner Isocenter Detector Rotation End Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

18

NUM

EV (128763, DCM, "Table Head Tilt End Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

19

NUM

EV (128764, DCM, "Table Horizontal Rotation End Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

20

NUM

EV (128765, DCM, "Table Cradle Tilt End Angle")

1

U

UNITS = EV (deg, UCUM, "deg")


Content Item Descriptions

Row 2

Angle in patient's "equatorial" plane (LAO to RAO). For dynamically changing angle during the event, the start value shall be provided. Equivalent to (0018,1510) in an image instance.

Row 3

Angle in patient's "sagittal" plane (CRAN to CAUD). For dynamically changing angle during the event, the start value shall be provided. Equivalent to (0018,1511) in an image instance.

Row 4

In case of motion during irradiation event, Positioner Primary ending angle

Row 5

In case of motion during irradiation event., Positioner Secondary ending angle

Row 6

Column device Angle in equipment based coordinates

Rows 12 to 20

Refer to the definitions of the X-Ray Isocenter Reference System (“X-Ray Isocenter Reference System Macro” in PS3.3 ).

DICOM PS3.16 2019a - Content Mapping Resource