DICOM PS3.16 2019c - Content Mapping Resource

Echocardiography Procedure Report Templates 

TID 5200 Echocardiography Procedure Report

This Template forms the top of a content tree that allows an ultrasound device to describe the results of an adult echocardiography imaging procedure. It is instantiated at the root node. It can also be included in other Templates that need to incorporate echocardiography findings into another report as quoted evidence.

Type:

Extensible

Order:

Significant

Root:

Yes

Table TID 5200. Echocardiography Procedure Report

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (125200, DCM, "Adult Echocardiography Procedure Report")

1

M

Root node

2

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

U

3

>

HAS OBS CONTEXT

INCLUDE

DTID 1001 “Observation Context”

1

M

4

>

CONTAINS

CONTAINER

DT (55111-9, LN, "Current Procedure Descriptions")

1

U

5

>>

CONTAINS

CODE

DT (125203, DCM, "Acquisition Protocol")

1-n

M

BCID 12001 “Ultrasound Protocol Types”

6

>

CONTAINS

INCLUDE

DTID 5201 “Echocardiography Patient Characteristics”

1

U

7

>

CONTAINS

CONTAINER

EV (111028, DCM, "Image Library")

1

U

8

>>

CONTAINS

IMAGE

1-n

M

9

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (87878005, SCT, "Left Ventricle")

$MeasType = DCID 12200 “Echocardiography Left Ventricle”

10

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (53085002, SCT, "Right Ventricle")

$MeasType = DCID 12204 “Echocardiography Right Ventricle”

11

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (82471001, SCT, "Left Atrium")

$MeasType = DCID 12205 “Echocardiography Left Atrium”

12

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (73829009, SCT, "Right Atrium")

$MeasType = DCID 12206 “Echocardiography Right Atrium”

13

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (34202007, SCT, "Aortic Valve")

$MeasType = DCID 12211 “Echocardiography Aortic Valve”

14

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (91134007, SCT, "Mitral Valve")

$MeasType = DCID 12207 “Echocardiography Mitral Valve”

15

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (39057004, SCT, "Pulmonic Valve")

$MeasType = DCID 12209 “Echocardiography Pulmonic Valve”

16

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (46030003, SCT, "Tricuspid Valve")

$MeasType = DCID 12208 “Echocardiography Tricuspid Valve”

17

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (15825003, SCT, "Aorta")

$MeasType = DCID 12212 “Echocardiography Aorta”

18

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (81040000, SCT, "Pulmonary artery")

$MeasType = DCID 12210 “Echocardiography Pulmonary Artery”

19

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (35532006, SCT, "Vena Cava")

$MeasType = DCID 12215 “Echocardiography Vena Cavae”

20

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (122972007, SCT, "Pulmonary Venous Structure")

$MeasType = DCID 12214 “Echocardiography Pulmonary Veins”

21

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (25489000, SCT, "Pericardial cavity")

$MeasType = DCID 12250 “Cardiac Ultrasound Common Linear Measurements”

22

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (241213007, SCT, "Cardiac Shunt Study")

$MeasType = DCID 12217 “Echocardiography Cardiac Shunt”

23

>

CONTAINS

INCLUDE

DTID 5202 “Echo Section”

1

U

$SectionSubject = EV (9904008, SCT, "Congenital Anomaly of Cardiovascular System")

$MeasType = DCID 12218 “Echocardiography Congenital”

24

>

CONTAINS

INCLUDE

DTID 5204 “Wall Motion Analysis”

1-n

U

$Procedure = DT (35757004, SCT, "Echocardiography for Determining Ventricular Contraction")


Content Item Descriptions

Row 8

No purpose of reference is specified.

Row 24

The wall motion findings of stress stage. There may be multiple Template instances to report wall motion findings of multiple stages.

DICOM PS3.16 2019c - Content Mapping Resource