DICOM PS3.16 2019a - Content Mapping Resource

CT Radiation Dose SR IOD Templates 

The Templates that comprise the CT Radiation Dose SR are interconnected as in Figure A-15.

CT Radiation Dose SR IOD Template Structure

Figure A-15. CT Radiation Dose SR IOD Template Structure


TID 10011 CT Radiation Dose

This Template defines a container (the root) with subsidiary Content Items, each of which corresponds to a single CT X-Ray irradiation event entry. There is a defined recording observer (the system or person responsible for recording the log, generally the system). Accumulated values shall be kept for a whole Study or at least a part of a Study, if the Study is divided in the workflow of the examination, or a performed procedure step. Multiple CT Radiation Dose objects may be created for one Study.

Type:

Extensible

Order:

Significant

Root:

Yes

Table TID 10011. CT Radiation Dose

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (113701, DCM, "X-Ray Radiation Dose Report")

1

M

Root node

1b

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

U

2

>

HAS CONCEPT MOD

CODE

EV (121058, DCM, "Procedure reported")

1

M

EV (P5-08000, SRT, "Computed Tomography X-Ray")

3

>>

HAS CONCEPT MOD

CODE

EV (G-C0E8, SRT, "Has Intent")

1

M

DCID 3629 “Procedure Intent”

4

>

INCLUDE

DTID 1002 “Observer Context”

1-n

M

5

>

HAS OBS CONTEXT

DATETIME

EV (113809, DCM, "Start of X-Ray Irradiation")

1

M

6

>

HAS OBS CONTEXT

DATETIME

EV (113810, DCM, "End of X-Ray Irradiation")

1

M

7

>

HAS OBS CONTEXT

CODE

EV (113705, DCM, "Scope of Accumulation")

1

M

DCID 10000 “Scope of Accumulation”

8

>>

HAS PROPERTIES

UIDREF

DCID 10001 “UID Types”

1

M

9

>

CONTAINS

INCLUDE

DTID 10012 “CT Accumulated Dose Data”

1

M

10

>

CONTAINS

INCLUDE

DTID 10013 “CT Irradiation Event Data”

1-n

M

11

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1

U

12

>

CONTAINS

CODE

EV (113854, DCM, "Source of Dose Information")

1-n

M

DCID 10021 “Source of CT Dose Information”

13

>

CONTAINS

INCLUDE

DTID 1020 “Person Participant”

1

U

$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing")


Content Item Descriptions

Row 4

The observer context may include both a Person Observer identification, as well as the identity of the equipment providing the values for the irradiation event (Device Observer identification), if not inherited.

Row 5

Start, DateTime of the first CT Irradiation Event of the accumulation

Row 6

End, DateTime of the last CT Irradiation Event of the accumulation

Row 12

The primary source of information from which this dose object was constructed.

Row 13

The physician responsible for determining that the irradiating procedure was appropriate for the indications. The value may come from Requesting Physician (0032,1032), Requesting Physician Identification Sequence (0032,1031) or somewhere else based on hospital policies.

DICOM PS3.16 2019a - Content Mapping Resource