DICOM PS3.7 2019d - Message Exchange

9.3.4.3 C-CANCEL-MOVE-RQ

The C-CANCEL-MOVE-RQ Message contains fields as defined in Table 9.3-11. Each field shall conform to DICOM encoding and Value Representation as defined in PS3.5. Fields are required as specified in the C-MOVE service definition unless otherwise noted in Table 9.3-11. Fields not specified in the C-MOVE service definition but present in Table 9.3-11 are required by the DIMSE-C protocol.

Table 9.3-11. C-CANCEL-MOVE-RQ Message Fields

Message Field

Tag

VR

VM

Description of Field

Command Group Length

(0000,0000)

UL

1

The even number of bytes from the end of the value field to the beginning of the next group.

Command Field

(0000,0100)

US

1

This field distinguishes the DIMSE-C operation conveyed by this Message. The value of this field shall be set to 0FFFH for the C-CANCEL-MOVE-RQ Message.

Message ID Being Responded To

(0000,0120)

US

1

Shall be set to the value of the Message ID (0000,0110) field used in associated C-MOVE-RQ Message.

Command Data Set Type

(0000,0800)

US

1

This field indicates that no Data Set is present in the Message and shall be set to a value of 0101H.


DICOM PS3.7 2019d - Message Exchange