DICOM PS3.7 2019a - Message Exchange

10.3.6.2 N-DELETE-RSP

The N-DELETE-RSP Message contains fields as defined in Table 10.3-12 and Annex C. Each field shall conform to DICOM encoding and Value Representation as defined in PS3.5. Fields are required as specified in the N-DELETE service definition unless otherwise noted in Table 10.3-12. Fields not specified in the N-DELETE service definition but present in Table 10.3-12 are required by the DIMSE-N protocol.

Table 10.3-12. N-DELETE-RSP Message Fields

Message Field

Tag

VR

VM

Description of Field

Command Group Length

(0000,0000)

UL

1

The even number of bytes from the end of the value field to the beginning of the next group.

Affected SOP Class UID

(0000,0002)

UI

1

SOP Class UID of the SOP Instance that was deleted.

Command Field

(0000,0100)

US

1

This field distinguishes the DIMSE-N operation conveyed by this Message. The value of this field shall be set to 8150H for the N-DELETE-RSP Message.

Message ID Being Responded To

(0000,0120)

US

1

Shall be set to the value of the Message ID (0000,0110) field used in associated N-DELETE-RQ Message.

Command Data Set Type

(0000,0800)

US

1

This field indicates that no Data Set is present in the Message. This field shall be set to the value of 0101H).

Status

(0000,0900)

US

1

The value of this field depends upon the status type. Annex C defines the encoding of the status types defined in the service definition.

Affected SOP Instance UID

(0000,1000)

UI

1

Contains the UID of the SOP Instance that was deleted.


DICOM PS3.7 2019a - Message Exchange