DICOM PS3.4 2019a - Service Class Specifications

F.9.4 Conformance Requirements

Implementations providing Standard SOP Class Conformance to the Modality Performed Procedure Step Notification SOP Class shall be conformant as described in the following sections and shall include within their Conformance Statement information as described in the following sections.

An implementation may conform to this SOP Class as an SCU, SCP or both. The Conformance Statement shall be in the format defined in PS3.2.

F.9.4.1 SCU Conformance

An implementation that is conformant to this SOP Class as an SCU shall meet conformance requirements for the:

  • notifications that it receives

F.9.4.1.1 Notifications

All standard event types for which notifications may be requested by the SCU shall be enumerated in the SCU Notifications Statement. The SCU Notifications Statement shall include an enumerated list of the event types supported:

  • Performed Procedure Step In Progress

  • Performed Procedure Step Completed

  • Performed Procedure Step Discontinued

  • Performed Procedure Step Updated

  • Performed Procedure Step Deleted

DICOM PS3.4 2019a - Service Class Specifications