DICOM PS3.4 2019a - Service Class Specifications

B.4.3 Conformance Statement Requirements

An implementation may conform to a SOP Class of the Storage Service Class as an SCU, SCP or both. The Conformance Statement shall be in the format defined in PS3.2.

B.4.3.1 Conformance Statement for an SCU

The following issues shall be documented in the Conformance Statement of any implementation claiming conformance to the Storage SOP Class as an SCU:

  • The behavior of the SCU in the case of a successful C-STORE response status shall be described.

  • The behavior of the SCU in each case of an unsuccessful C-STORE response status shall be described.

  • The behavior of the SCU in the case of a Warning status received in response to a C-STORE operation.

  • Whether extended negotiation is supported.

  • The optional elements that may be included in Storage SOP Instances for each IOD supported shall be listed.

  • The standard and privately defined Functional Groups that may be included in Storage SOP Instances for each Multi-frame IOD that support Functional Groups.

  • The behavior of the SCU in the case of a C-STORE operation using a referenced pixel data transfer syntax such as JPIP Referenced Pixel Data Transfer Syntax shall be described. This includes the duration of validity of the reference

DICOM PS3.4 2019a - Service Class Specifications