DICOM PS3.4 2020e - Service Class Specifications

B.2.3 Statuses

Table B.2-1 defines the specific status code values that might be returned in a C-STORE response. General status code values and fields related to status code values are defined for C-STORE DIMSE Service in PS3.7.

Table B.2-1. C-STORE Status

Service Status

Further Meaning

Status Codes

Related Fields

Failure

Refused: Out of resources

A7xx

(0000,0902)

Error: Data Set does not match SOP Class

A9xx

(0000,0901)

(0000,0902)

Error: Cannot understand

Cxxx

(0000,0901)

(0000,0902)

Warning

Coercion of Data Elements

B000

(0000,0901)

(0000,0902)

Data Set does not match SOP Class

B007

(0000,0901)

(0000,0902)

Elements Discarded

B006

(0000,0901)

(0000,0902)

Success

Success

0000

None


Some Failure Status Codes are implementation specific.

An SCP implementation shall assign specific failure status codes by replacing each 'x' symbol with a hexadecimal digit in the range from 0 to F. An SCP implementation wishing to differentiate between causes within the same Failure Meaning shall assign those causes specific Status Code Values within valid range specified in Table B.2-1.

An SCU implementation shall recognize any Failure Status Code within the value range specified in Table B.2-1 as an indicator of the Failure Meaning stated in the table. There is no requirement for an SCU implementation to differentiate between specific Status Codes within the valid range.

DICOM PS3.4 2020e - Service Class Specifications