DICOM PS3.3 2020e - Information Object Definitions |
---|
This Module describes the Attributes of a Tomography acquisition (translation of X-Ray source during the acquisition of a single-frame image).
Table C.8-32. X-Ray Tomography Acquisition Module Attributes
Distance in mm between the table surface and the sharp image plane. |
|||
Angle span in degrees of rotation of X-Ray Source during X-Ray acquisition. |
|||
Time in seconds the source has taken to rotate the Tomo Angle during X-Ray acquisition. |
|||
The number of source images used to construct this tomosynthetic image. Only meaningful if Tomo Class (0018,1491) is TOMOSYNTHESIS. These may be listed in Source Image Sequence (0008,2112) of the General Image Module. |
DICOM PS3.3 2020e - Information Object Definitions |
---|