DICOM PS3.3 2021d - Information Object Definitions

C.8.6.4 SC Multi-frame Vector Module

Table C.8-25c contains IOD Attributes that may be the target of the Frame Increment Pointer (0028,0009) for SC Multi-frame images.

Table C.8-25c. SC Multi-frame Vector Module Attributes

Attribute Name

Tag

Type

Attribute Description

Frame Time Vector

(0018,1065)

1C

An array that contains the real time increments (in msec) between frames for a Multi-frame image.

See Section C.7.6.5.1.2 for further explanation.

Required if Frame Increment Pointer (0028,0009) points to Frame Time Vector (0018,1065).

Note

Frame Time Vector arrays may not be properly encoded if Explicit VR Transfer Syntax is used and the VL of this Attribute exceeds 65534 bytes.

Page Number Vector

(0018,2001)

1C

An array that contains, for each of the image frames, the corresponding page numbers of the original document.

Required if Frame Increment Pointer (0028,0009) points to Page Number Vector (0018,2001).

Frame Label Vector

(0018,2002)

1C

An array that contains, for each of the image frames, a descriptive label.

Required if Frame Increment Pointer (0028,0009) points to Frame Label Vector (0018,2002).

Frame Primary Angle Vector

(0018,2003)

1C

An array that contains, for each of the image frames, the primary angle of rotation about an undefined axis, in degrees. May be used for annotative purposes for "cine loops" of 3D reprojected images

Required if Frame Increment Pointer (0028,0009) points to Frame Primary Angle Vector (0018,2003).

Frame Secondary Angle Vector

(0018,2004)

1C

An array that contains, for each of the image frames, the secondary angle of rotation about an undefined axis that is orthogonal to that used for Frame Primary Angle Vector (0018,2003), in degrees. May be used for annotative purposes for "cine loops" of 3D reprojected images

Required if Frame Increment Pointer (0028,0009) points to Frame Secondary Angle Vector (0018,2004).

Slice Location Vector

(0018,2005)

1C

An array that contains, for each of the image frames, the relative position of the image plane expressed in mm, as defined for Slice Location (0020,1041). See Section C.7.6.2.1.2 for further explanation.

Required if Frame Increment Pointer (0028,0009) points to Slice Location Vector (0018,2005).

Display Window Label Vector

(0018,2006)

1C

An array that contains, for each of the image frames, a label or number of the display window of a graphical user interface from which the frame was captured.

Required if Frame Increment Pointer (0028,0009) points to Display Window Label Vector (0018,2006).


DICOM PS3.3 2021d - Information Object Definitions