DICOM PS3.3 2020e - Information Object Definitions |
---|
This section describes the specific Modules for the Surface Segmentation IOD.
Table C.8.23-1 defines the general Attributes of the Surface Segmentation Module.
Table C.8.23-1. Surface Segmentation Module Attributes
Include Table 10-12 “Content Identification Macro Attributes” |
|||
>Include Table C.8.20-4 “Segment Description Macro Attributes” |
|||
The number of surfaces that comprise this segment. Shall be greater than zero. |
|||
Sequence referencing the surfaces composed to construct this segment. One or more Items shall be included in this Sequence. The number of Items shall equal the value of Surface Count (0066,002A). |
|||
Identifies the Surface Number (0066,0003) within the Surface Sequence (0066,0002) to which this reference applies. |
|||
>>Segment Surface Generation Algorithm Identification Sequence |
|||
>>>Include Table 10-19 “Algorithm Identification Macro Attributes” |
For Algorithm Family Code Sequence (0066,002F) BCID 7162 “Surface Processing Algorithm Families”. |
||
A Sequence that identifies the set of Instances by their SOP Class/Instance pair that were used to derive this segment surface. |
|||
>>>Include Table 10-3 “Image SOP Instance Reference Macro Attributes” |
DICOM PS3.3 2020e - Information Object Definitions |
---|