Radiopharmaceutical Information Sequence
|
(0054,0016)
|
1
|
Sequence of Items that describe isotope information.
One or more Items shall be included in this Sequence.
|
>Radiopharmaceutical Agent Number
|
(0018,9729)
|
1
|
Identifying number, unique within this SOP Instance, of the agent administered. Used to reference this particular agent from the Radiopharmaceutical Functional Group Macro. The number shall be 1 for the first Item and increase by 1 for each subsequent Item.
|
>Radionuclide Code Sequence
|
(0054,0300)
|
1
|
Sequence that identifies the radionuclide.
Only a single Item shall be included in this Sequence.
|
>>Include Table 8.8-1 “Code Sequence Macro Attributes”
|
BCID 4020 “PET Radionuclide”.
|
>Administration Route Code Sequence
|
(0054,0302)
|
1
|
Sequence that identifies the administration route of the radiopharmaceutical.
Only a single Item shall be included in this Sequence.
|
>>Include Table 8.8-1 “Code Sequence Macro Attributes”
|
BCID 11 “Route of Administration”.
|
>Radiopharmaceutical Volume
|
(0018,1071)
|
3
|
Volume of administered radiopharmaceutical in cubic cm.
|
>Radiopharmaceutical Start DateTime
|
(0018,1078)
|
1
|
Time of start of administration. The actual time of radiopharmaceutical administration to the patient for imaging purposes.
|
>Radiopharmaceutical Stop DateTime
|
(0018,1079)
|
3
|
Time of end of administration. The actual ending time of radiopharmaceutical administration to the patient for imaging purposes.
|
>Radionuclide Total Dose
|
(0018,1074)
|
2
|
The radiopharmaceutical dose administered to the patient measured in MegaBecquerels (MBq) at the Radiopharmaceutical Start DateTime (0018,1078).
|
>Radiopharmaceutical Administration Event UID
|
(0008,3012)
|
3
|
Unique identification of the administration of the radiopharmaceutical to the patient.
Note
The UID is the same Radiopharmaceutical Administration Event UID that is in the Radiopharmaceutical Radiation Dose Report.
|
>Radionuclide Half Life
|
(0018,1075)
|
1
|
The radionuclide half life, in seconds, that was used in the correction of this image.
|
>Radionuclide Positron Fraction
|
(0018,1076)
|
1
|
The radionuclide positron fraction (fraction of decays that are by positron emission) that was used in the correction of this image.
|
>Radiopharmaceutical Specific Activity
|
(0018,1077)
|
3
|
The activity per unit mass of the radiopharmaceutical, in Bq/micromole, at the Radiopharmaceutical Start DateTime (0018,1078).
|
>Radiopharmaceutical Code Sequence
|
(0054,0304)
|
1
|
Sequence that identifies the radiopharmaceutical.
Only a single Item shall be included in this Sequence.
|
>>Include Table 8.8-1 “Code Sequence Macro Attributes”
|
BCID 4021 “PET Radiopharmaceutical”.
|