DICOM PS3.3 2021a - Information Object Definitions |
---|
Table C.8.20-2 defines the general Attributes of the Segmentation Image Module.
Table C.8.20-2. Segmentation Image Module Attributes
Value 1 shall be DERIVED. Value 2 shall be PRIMARY. No other values shall be present. |
|||
Include Table 10-12 “Content Identification Macro Attributes” |
|||
Total number of columns in pixel matrix; i.e., width of total imaged volume in pixels. See Section C.8.12.4.1.3 Required if Dimension Organization Type (0020,9311) is present with a value of TILED_FULL. May be present otherwise. |
|||
Total number of rows in pixel matrix; i.e., height of total imaged volume in pixels. See Section C.8.12.4.1.3 Required if Dimension Organization Type (0020,9311) is present with a value of TILED_FULL. May be present otherwise. |
|||
Total number of focal planes (Z locations) in the pixel matrix; i.e., depth of total imaged volume in pixels. See Section C.8.12.4.1.3 Required if Dimension Organization Type (0020,9311) is present with a value of TILED_FULL. May be present otherwise. |
|||
Location of pixel 1\1 of the total pixel matrix in the Slide Coordinate System Frame of Reference. Only a single Item shall be included in this Sequence. See Section C.8.12.4.1.4 and Section C.8.12.2.1.1 for further explanation. Required if Dimension Organization Type (0020,9311) is present with a value of TILED_FULL. May be present otherwise. |
|||
The X offset in millimeters from the Origin of the Slide Coordinate System. |
|||
The Y offset in millimeters from the Origin of the Slide Coordinate System. |
|||
The direction cosines of the first row and the first column of the total pixel matrix with respect to the Slide Coordinate System Frame of Reference. See Section C.8.12.4.1.4. Required if Plane Position (Slide) Sequence (0048,021A) is present within a Functional Group Sequence or Dimension Organization Type (0020,9311) is present with a value of TILED_FULL. May be present otherwise. |
|||
See Section C.8.20.2.1. |
|||
See Section C.8.20.2.1. |
|||
See Section C.8.20.2.1. |
|||
Specifies whether an Image has undergone lossy compression (at a point in its lifetime), or is derived from lossy compressed images. Enumerated Values: Once this value has been set to 01 it shall not be reset. See Section C.8.20.2.2 and Section C.7.6.1.1.5, “Lossy Image Compression”. |
|||
Describes the approximate lossy compression ratio(s) that have been applied to this image. Required if present in the source images or this IOD Instance has been compressed. |
|||
A label for the lossy compression method(s) that have been applied to this image. Required if present in the source images or this IOD Instance has been compressed. See Section C.8.20.2.2. |
|||
The type of encoding used to indicate the presence of the segmented property at a pixel/voxel location. See Section C.8.20.2.3. |
|||
For fractional segmentation encoding, the meaning of the fractional value. Required if Segmentation Type (0062,0001) is FRACTIONAL. See Section C.8.20.2.3 for Enumerated Values. |
|||
Specifies the value that represents a probability of 1 or complete occupancy. There shall be no values in Pixel Data (7FE0,0010) greater than this value. Required if Segmentation Type (0062,0001) is FRACTIONAL. |
|||
Whether or not any segments in this Instance overlap. I.e., whether or not any pixel is or might be in more than one segment. Enumerated Values: See Section C.8.20.2.3. |
|||
>Include Table C.8.20-4 “Segment Description Macro Attributes” |
|||
Name of algorithm used to generate the segment. Required if Segment Algorithm Type (0062,0008) is not MANUAL. |
|||
A description of how this segment was derived. Algorithm Name (0066,0036) within this Sequence may be identical to Segment Algorithm Name (0062,0009). |
|||
>>Include Table 10-19 “Algorithm Identification Macro Attributes” |
|||
A default single gray unsigned value in which it is recommended that the maximum pixel value in this segment be rendered on a monochrome display. The units are specified in P-Values from a minimum of 0000H (black) up to a maximum of FFFFH (white). |
|||
A default triplet value in which it is recommended that segment be rendered on a color display. The units are specified in PCS-Values, and the value is encoded as CIELab. See Section C.10.7.1.1. |
As a consequence of the enumerated Bits Allocated and Bits Stored Attribute values, single bit pixels shall be packed 8 to a byte as defined by the encoding rules in PS3.5.
DICOM PS3.3 2021a - Information Object Definitions |
---|