DICOM PS3.3 2021a - Information Object Definitions |
---|
This Section describes the CT Image Module. This Module contains all Attributes that are specific to CT images.
The table in this Section contains IOD Attributes that describe CT images.
Table C.8-3. CT Image Module Attributes
Image identification characteristics. See Section C.8.2.1.1.1 for specialization. |
|||
Indicates whether the image is created by means of Multi-energy technique. |
|||
Number of samples (planes) in this image. See Section C.8.2.1.1.2 for specialization. |
|||
Specifies the intended interpretation of the pixel data. See Section C.8.2.1.1.3 for specialization. |
|||
Number of bits allocated for each pixel sample. Each sample shall have the same number of bits allocated. See Section C.8.2.1.1.4 for specialization. |
|||
Number of bits stored for each pixel sample. Each sample shall have the same number of bits stored. See Section C.8.2.1.1.5 for specialization. |
|||
Most significant bit for pixel sample data. Each sample shall have the same high bit. See Section C.8.2.1.1.6 for specialization. |
|||
The value b in relationship between stored values (SV) and the output units. If Image Type (0008,0008) Value 1 is ORIGINAL and Value 3 is not LOCALIZER, and Multi-energy CT Acquisition (0018,9361) is either absent or NO, output units shall be Hounsfield Units (HU). |
|||
m in the equation specified in Rescale Intercept (0028,1052). |
|||
Specifies the output units of Rescale Slope (0028,1053) and Rescale Intercept (0028,1052). See Section C.11.1.1.2 for Defined Terms and further explanation. Required if the Rescale Type is not HU (Hounsfield Units), or Multi-energy CT Acquisition (0018,9361) is YES. May be present otherwise. |
|||
Peak kilo voltage output of the X-Ray generator used. Shall be empty if this Attribute is present in Multi-energy CT Acquisition Sequence (0018,9362). |
|||
A number identifying the single continuous gathering of data over a period of time that resulted in this image |
|||
The diameter in mm of the region over which data were collected. Shall not be present if this Attribute is present in Multi-energy CT Acquisition Sequence (0018,9362) and the value of this Attribute is not the same in all Items of the Multi-energy CT Acquisition Sequence (0018,9362). |
|||
The x, y, and z coordinates (in the Patient-Based Coordinate System) in mm of the center of the region in which data were collected. See Section C.8.15.3.6.1. |
|||
Diameter in mm of the region from within which data were used in creating the reconstruction of the image. Data may exist outside this region and portions of the patient may exist outside this region. The diameter defines a circular region that is entirely contained within the encoded Pixel Data (7FE0,0010), unless the encoded image has been cropped after reconstruction. |
|||
The x, y, and z coordinates (in the Patient-Based Coordinate System) of the reconstruction center target point as used for reconstruction in mm. See Section C.8.15.3.6.1. |
|||
Distance in mm from source to detector center. NoteThis value is traditionally referred to as Source Image Receptor Distance (SID). Shall not be present if this Attribute is present in Multi-energy CT Acquisition Sequence (0018,9362) and the value of this Attribute is not the same in all Items of the Multi-energy CT Acquisition Sequence (0018,9362). |
|||
Distance in mm from source to isocenter (center of field of view). |
|||
Nominal angle of tilt in degrees of the scanning gantry. Not intended for mathematical computations. |
|||
The distance in mm of the top of the patient table to the center of rotation; below the center is positive. |
|||
Direction of rotation of the source when relevant, about nearest principal axis of equipment. |
|||
Time of X-Ray exposure in msec. If Acquisition Type (0018,9302) equals SPIRAL, the value of this Attribute shall be Revolution Time (0018,9305) divided by the Spiral Pitch Factor (0018,9311). See Section C.8.15.3.8.1 and Section C.8.15.3.2.1. Shall not be present if the corresponding Attribute, Exposure Time in ms (0018,9328), is present in Multi-energy CT Acquisition Sequence (0018,9362) and the value of this Attribute is not the same in all Items of the Multi-energy CT Acquisition Sequence (0018,9362). |
|||
Shall not be present if the corresponding Attribute, X-Ray Tube Current in mA (0018,9330), is present in Multi-energy CT Acquisition Sequence (0018,9362) and the value of this Attribute is not the same in all Items of the Multi-energy CT Acquisition Sequence (0018,9362). |
|||
The exposure expressed in mAs, for example calculated from Exposure Time and X-Ray Tube Current. Shall not be present if the corresponding Attribute, Exposure in mAs (0018,9332), is present in Multi-energy CT Acquisition Sequence (0018,9362) and the value of this Attribute is not the same in all Items of the Multi-energy CT Acquisition Sequence (0018,9362). |
|||
The exposure expressed in µAs, for example calculated from Exposure Time and X-Ray Tube Current. Shall not be present if the corresponding Attribute, Exposure in mAs (0018,9332), is present in Multi-energy CT Acquisition Sequence (0018,9362) and the value of this Attribute is not the same in all Items of the Multi-energy CT Acquisition Sequence (0018,9362). |
|||
Label for the type of filter inserted into the X-Ray beam. Shall not be present if this Attribute is present in Multi-energy CT Acquisition Sequence (0018,9362) and the value of this Attribute is not the same in all Items of the Multi-energy CT Acquisition Sequence (0018,9362). |
|||
Power in kW to the X-Ray generator. Shall not be present if this Attribute is present in Multi-energy CT Acquisition Sequence (0018,9362) and the value of this Attribute is not the same in all Items of the Multi-energy CT Acquisition Sequence (0018,9362). |
|||
Size of the focal spot in mm. For devices with variable focal spot or multiple focal spots, small dimension followed by large dimension. Shall not be present if this Attribute is present in Multi-energy CT Acquisition Sequence (0018,9362) and the value of this Attribute is not the same in all Items of the Multi-energy CT Acquisition Sequence (0018,9362). |
|||
A label describing the convolution kernel or algorithm used to reconstruct the data |
|||
The time in seconds of a complete revolution of the source around the gantry orbit. |
|||
The width of a single row of acquired data (in mm). NoteAdjacent physical detector rows may have been combined to form a single effective acquisition row. Shall not be present if this Attribute is present in Multi-energy CT Acquisition Sequence (0018,9362) and the value of this Attribute is not the same in all Items of the Multi-energy CT Acquisition Sequence (0018,9362). |
|||
The width of the total collimation (in mm) over the area of active X-Ray detection. NoteThis will be equal the number of effective detector rows multiplied by single collimation width. Shall not be present if this Attribute is present in Multi-energy CT Acquisition Sequence (0018,9362) and the value of this Attribute is not the same in all Items of the Multi-energy CT Acquisition Sequence (0018,9362). |
|||
The distance in mm that the table moves in one second during the gathering of data that resulted in this image. |
|||
Motion of the table (in mm) during a complete revolution of the source around the gantry orbit. |
|||
Ratio of the Table Feed per Rotation (0018,9310) to the Total Collimation Width (0018,9307). |
|||
A label describing the type of exposure modulation used for the purpose of limiting the dose. |
|||
A percent value of dose saving due to the use of Exposure Modulation Type (0018,9323). A negative percent value of dose savings reflects an increase of exposure. |
|||
Computed Tomography Dose Index (CTDIvol), in mGy according to IEC 60601-2-44, Ed.2.1 (Clause 29.1.103.4), The Volume CTDIvol. It describes the average dose for this image for the selected CT conditions of operation. |
|||
The type of phantom used for CTDI measurement according to [IEC 60601-2-44]. |
|||
The diameter, in mm, of a cylinder of water having the same X-Ray attenuation as the patient for this reconstructed slice (e.g., as described in [AAPM Report 220]). |
|||
The method of calculation of Water Equivalent Diameter (0018,1271). Required if Water Equivalent Diameter (0018,1271) is present. |
|||
X-Ray dose, measured in dGy*cm*cm, to which the patient was exposed for the acquisition of the entire Irradiation Event from which this image was reconstructed. Note
|
|||
Include Table 10-7 “General Anatomy Optional Macro Attributes” |
Anatomic Region Sequence DCID 4030 “CT, MR and PET Anatomy Imaged”. |
||
Include Table 10-25 “Optional View and Slice Progression Direction Macro Attributes” |
|||
The calibration factor for the calcium mass score. These factors incorporate the effects of See Section C.8.2.1.1.7. |
|||
The calibration factors for the calcium mass score of the device. These factors incorporate the effects of This a multi-value Attribute, the first value specifies the mass factor for a small patient size, the second value for a medium patient size and the third value for a large patient size. See Section C.8.2.1.1.7. |
|||
The weighting factor of the data from the primary source in a multiple energy composition image. This factor incorporates the effects of Required if one Derivation Code Sequence (0008,9215) Item value is (113097, DCM, "Multi-energy proportional weighting"). May be present otherwise. |
|||
Acquisition parameters for X-Ray sources beyond the primary X-Ray source, which is specified in other Attributes of this Module. One or more Items are permitted in this Sequence. NoteThis sequence is superseded for multi-energy acquisitions by the Multi-energy CT Acquisition Sequence (0018,9362), however implementations may encounter instances that still use this sequence, in which case Multi-energy CT Acquisition (0018,9361) will be absent or will be set to NO. Shall not be present if Multi-energy CT Acquisition (0018,9361) is YES. |
|||
The diameter in mm of the region over which data were collected. |
|||
The exposure expressed in milliampere seconds, for example calculated from exposure time and X-Ray tube current. |
|||
The weighting factor of the data from this additional source in a multiple energy composition image. This factor incorporates the effects of Required if one Derivation Code Sequence (0008,9215) Item value is (113097, DCM, "Multi-energy proportional weighting"). May be present otherwise. |
|||
Isocenter coordinates (x,y,z), in mm. Specifies the location of the machine isocenter in the Patient-Based Coordinate System associated with the Frame of Reference. It allows transformation from the Equipment-Based Coordinate System to the Patient-Based Coordinate System. |
|||
Include Table 10-27 “RT Equipment Correlation Macro Attributes Description” |
For CT Images, Image Type (0008,0008) is specified to be Type 1.
Image Type (0008,0008) Value 4 shall be present if Multi-energy CT Acquisition (0018,9361) has a value of YES.
Defined Terms for Value 4 for Multi-energy CT Images:
a Virtual Monoenergetic Image. Each real-world value mapped pixel represents CT Hounsfield units and is analogous to a CT image created by a monoenergetic (of a specific keV value) X-Ray beam.
a Material-Specific Image. Each real-world value mapped pixel value represents a property of a material such as attenuation, concentration or density.
An image with the attenuation contribution of one or more materials removed. For pixels that did not contain any of the removed material(s), the pixel values are unchanged.
a Material-Fractional Image. Each real-world value mapped pixel represents the fraction of a voxel occupied by a material.
an Effective Atomic Number Image. Each real-world value mapped pixel represents Effective Atomic Number of the materials in the voxel.
an Electron Density Image. Each real-world value mapped pixel represents the number of electrons per unit volume or the electron density relative to water.
a Material-Modified Image. CT Image where real-world value mapped pixels have been modified to highlight a certain target material (either by partially suppressing the background or by enhancing the target material), or to partially suppress the target material.
a Value-Based Image. CT Image where real-world value mapped pixels represent a certain value for a specified material
When an image is created by a generic transformation an implementation-specific Value 4 may be provided.
See Section C.7.6.3.1.2 for definition of these terms.
The calibration factors for the Calcium Scoring Mass Factor Patient (0018,9351) and Calcium Scoring Mass Factor Device (0018,9352) Attributes are defined by the International Consortium for Multi-Detector CT Evaluation of Coronary Calcium, see McCollough, C.H. "A multi-institutional, multi-manufacturer, international standard for the quantification of coronary artery calcium using cardiac CT".
DICOM PS3.3 2021a - Information Object Definitions |
---|