DICOM PS3.3 2020a - Information Object Definitions
C.8.19.6.5 X-Ray Frame Detector Parameters Macro

Table C.8.19.6-5 specifies the Attributes containing the X-Ray Frame Detector Parameters Functional Group Macro.

Table C.8.19.6-5. X-Ray Frame Detector Parameters Macro Attributes

Attribute Name

Tag

Type

Attribute Description

Frame Detector Parameters Sequence

(0018,9451)

1

Sequence containing the detector properties for this frame.

Only a single Item shall be included in this Sequence.

>Detector Active Time

(0018,7014)

3

Time in mSec that the detector is active during acquisition of this image.

Note

This activation window overlaps the time of the X-Ray exposure as defined by Exposure Time in ms (0018,9328) and Detector Activation Offset From Exposure (0018,7016).

>Detector Activation Offset From Exposure

(0018,7016)

3

Offset time in mSec that the detector becomes active after the X-Ray beam is turned on during acquisition of this image. May be negative.


DICOM PS3.3 2020a - Information Object Definitions