DICOM PS3.3 2021a - Information Object Definitions

C.8.19.4 X-Ray Image Intensifier Module

Table C.8.19.4-1 specifies the Attributes of the X-Ray Image Intensifier Module.

Table C.8.19.4-1. X-Ray Image Intensifier Module Attributes

Attribute Name

Tag

Type

Attribute Description

Intensifier Size

(0018,1162)

1

Physical diameter of the maximum active area X-Ray intensifier in mm.

Note

This Attribute does not specify the field of view. The Attribute Field of View Dimension(s) in Float (0018,9461) is intended for this value.

Intensifier Active Shape

(0018,9427)

1

Shape of the active area used for acquiring this image.

Enumerated Values:

RECTANGLE

ROUND

HEXAGONAL

Note

This may be different from the Field of View Shape (0018,1147), and should not be assumed to describe the stored image.

Intensifier Active Dimension(s)

(0018,9428)

1

Dimensions in mm of the active area used for acquiring this image.

If Intensifier Active Shape (0018,9427) is:

RECTANGLE: row dimension followed by column.

ROUND: diameter.

HEXAGONAL: diameter of the circle circumscribing the hexagon.

Note

This may be different from the Field of View Dimension(s) in Float (0018,9461), and should not be assumed to describe the stored image.


DICOM PS3.3 2021a - Information Object Definitions