DICOM PS3.3 2021d - Information Object Definitions

C.8.15 Enhanced CT Image

This section describes the specific Modules for the Enhanced CT Image IOD.

C.8.15.1 CT Series Module

The CT IODs use the General Series Module, specialized by the CT Series Module, to describe the DICOM Series Entity described in Section A.1.2.3, and to define what constitutes a Series for the context of CT device.

Table C.8-113 specifies the Attributes that identify and describe general information about the CT Series.

Table C.8-113. CT Series Module Attributes

Attribute Name

Tag

Type

Attribute Description

Modality

(0008,0060)

1

Type of device, process or method that originally acquired the data used to create the images in this Series.

Enumerated Values:

CT

See Section C.7.3.1.1.1 for further explanation.

Referenced Performed Procedure Step Sequence

(0008,1111)

1C

Uniquely identifies the Performed Procedure Step SOP Instance to which the Series is related.

Only a single Item shall be included in this Sequence.

Required if a Performed Procedure Step SOP Class was involved in the creation of this Series.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”


DICOM PS3.3 2021d - Information Object Definitions