DICOM PS3.3 2021d - Information Object Definitions
C.8.13.5.8 MR Transmit Coil Macro

Table C.8-95 specifies the Attributes of the MR Transmit Coil Functional Group Macro.

Table C.8-95. MR Transmit Coil Macro Attributes

Attribute Name

Tag

Type

Attribute Description

MR Transmit Coil Sequence

(0018,9049)

1

A Sequence that provides information about the transmit coil used.

Only a single Item shall be included in this Sequence.

>Transmit Coil Name

(0018,1251)

1C

Name of transmit coil used.

Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL. May be present otherwise.

>Transmit Coil Manufacturer Name

(0018,9050)

2C

Name of manufacturer of transmit coil.

Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL. May be present otherwise.

>Transmit Coil Type

(0018,9051)

1C

Type of transmit coil used.

Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL. May be present otherwise.

Defined Terms:

BODY

VOLUME

head, extremity, etc.

SURFACE


DICOM PS3.3 2021d - Information Object Definitions