DICOM PS3.3 2019c - Information Object Definitions

C.3.4 Visit Admission Module

Table C.3-4 defines the Attributes relevant to admitting a Patient during a Visit.

Table C.3-4. Visit Admission Module Attributes

Attribute Name

Tag

Attribute Description

Referring Physician's Name

(0008,0090)

Patient's primary referring physician for this Visit

Referring Physician's Address

(0008,0092)

Referring physician's address

Referring Physician's Telephone Numbers

(0008,0094)

Referring physician's telephone numbers

Referring Physician Identification Sequence

(0008,0096)

Identification of the Patient's referring physician. Only a single Item shall be included in this Sequence.

>Include Table 10-1 “Person Identification Macro Attributes Description”

Note

The Person's Telecom Information (0040,1104) Attribute in this Sequence may be treated as equivalent to HL7v2 field ROL-12 associated with the referring physician identified in PV1-8.

Consulting Physician's Name

(0008,009C)

Consulting physician(s) for this Patient Visit.

Consulting Physician Identification Sequence

(0008,009D)

Identification of the consulting physician(s).

One or more Items shall be included in this Sequence. If more than one Item, the number and order shall correspond to the value of Consulting Physician's Name (0008,009C), if present.

>Include Table 10-1 “Person Identification Macro Attributes Description”

Note

The Person's Telecom Information (0040,1104) Attribute in this Sequence may be treated as equivalent to HL7v2 field ROL-12 associated with the consulting physician.

Admitting Diagnoses Description

(0008,1080)

Description of admitting diagnosis (diagnoses).

Admitting Diagnoses Code Sequence

(0008,1084)

A Sequence that conveys the admitting diagnosis (diagnoses). One or more Items shall be included in this Sequence.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

No Baseline CID is defined.

Route of Admissions

(0038,0016)

Mode of admission: emergency, normal

Admitting Date

(0038,0020)

Date Patient Visit began

Admitting Time

(0038,0021)

Time Patient Visit began


DICOM PS3.3 2019c - Information Object Definitions