DICOM PS3.3 2020b - Information Object Definitions

C.25.2 Real World Value Mapping Module

Table C.25.2-1 defines the general Attributes of the Real World Value Mapping Module.

Table C.25.2-1. Real World Value Mapping Module Attributes

Attribute Name

Tag

Type

Attribute Description

Content Date

(0008,0023)

1

The date the content creation started.

Content Time

(0008,0033)

1

The time the content creation started.

Include Table 10-12 “Content Identification Macro Attributes”

Referenced Image Real World Value Mapping Sequence

(0040,9094)

1

A Sequence of real world value mapping Items. Each Item defines a single mapping and a list of images to which the mapping applies.

One or more Items shall be included in this Sequence.

>Include Table C.7.6.16-12 “Real World Value Mapping Macro Attributes”

Defined CID for Measurement Units Code Sequence is CID 83 “Units for Real World Value Mapping”.

>Referenced Image Sequence

(0008,1140)

1

A Sequence listing the images to which the mapping applies.

One or more Items shall be included in this Sequence.

>>Include Table 10-3 “Image SOP Instance Reference Macro Attributes”


DICOM PS3.3 2020b - Information Object Definitions