DICOM PS3.3 2021b - Information Object Definitions

C.11.25 Presentation View Description Module

Table C.11.25-1 contains Attributes that describe the view of this presentation in anatomical terms.

Note: For example, a Volumetric Presentation State derived from a whole-body CT source volume could be restricted to the left leg.

Table C.11.25-1. Presentation View Description Module Attributes

Attribute Name

Tag

Type

Attribute Description

Include Table 10-6 “General Anatomy Required Macro Attributes”

No Baseline CID is defined.

View Code Sequence

(0054,0220)

2

Describes this view of the patient anatomy.

No more than one Item shall be included in this Sequence.

>Include Table 8.8-1 “Code Sequence Macro Attributes” .

BCID 501 “Volumetric View Description”.

>View Modifier Code Sequence

(0054,0222)

2

View modifier.

Zero or more Items shall be included in this Sequence.

>> Include Table 8.8-1 “Code Sequence Macro Attributes” .

BCID 502 “Volumetric View Modifier”.

Image Laterality

(0020,0062)

2

Laterality of (possibly paired) body part (as described in Anatomic Region Sequence (0008,2218) ) examined.

Enumerated Values:

R

right

L

left

U

unpaired

B

both left and right

Shall be consistent with any laterality information contained in Primary Anatomic Structure Modifier Sequence (0008,2230), if present.


DICOM PS3.3 2021b - Information Object Definitions