DICOM PS3.3 2020b - Information Object Definitions

A.36.3.3 MR Spectroscopy IOD Module Table

Table A.36-3 specifies the Modules of the MR Spectroscopy IOD.

Table A.36-3. MR Spectroscopy IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

General Series

C.7.3.1

M

Clinical Trial Series

C.7.3.2

U

MR Series

C.8.13.6

M

Frame of Reference

Frame of Reference

C.7.4.1

M

Synchronization

C.7.4.2

C - Required if time synchronization was applied.

Equipment

General Equipment

C.7.5.1

M

Enhanced General Equipment

C.7.5.2

M

MR Spectroscopy

Enhanced Contrast/Bolus

C.7.6.4b

C - Required if contrast media were applied.

Multi-frame Functional Groups

C.7.6.16

M

Multi-frame Dimension

C.7.6.17

M

Cardiac Synchronization

C.7.6.18.1

C - Required if cardiac synchronization was applied.

Respiratory Synchronization

C.7.6.18.2

C - Required if respiratory synchronization was applied.

Bulk Motion Synchronization

C.7.6.18.3

C - Required if bulk motion synchronization was applied.

Acquisition Context

C.7.6.14

M

Specimen

C.7.6.22

U

MR Spectroscopy

C.8.14.1

M

MR Spectroscopy Pulse Sequence

C.8.14.2

C - Required if Image Type (0008,0008) Value 1 is ORIGINAL. May be present otherwise.

MR Spectroscopy Data

C.8.14.4

M

SOP Common

C.12.1

M

Common Instance Reference

C.12.2

U

Frame Extraction

C.12.3

C - Required if the SOP Instance was created in response to a Frame-Level retrieve request


Note

The Specimen Identification Module was previously included in this IOD but has been retired, and its functionality replaced by the Specimen Module. See PS3.3-2008.

DICOM PS3.3 2020b - Information Object Definitions