DICOM PS3.3 2020b - Information Object Definitions

A.35.17.3 Simplified Adult Echo SR IOD Module Table

Table A.35.17-1 specifies the Modules of the Simplified Adult Echo SR IOD.

Table A.35.17-1. Simplified Adult Echo SR IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

SR Document Series

C.17.1

M

Clinical Trial Series

C.7.3.2

U

Frame of Reference

Synchronization

C.7.4.2

C - Required if time synchronization was applied

Equipment

General Equipment

C.7.5.1

M

Enhanced General Equipment

C.7.5.2

M

Document

SR Document General

C.17.2

M

SR Document Content

C.17.3

M

Timezone

C.12.5

M

SOP Common

C.12.1

M


A.35.17.3.1 Simplified Adult Echo SR IOD Content Constraints
A.35.17.3.1.1 Template

The document shall be constructed from TID 5300 “Simplified Echo Procedure Report” invoked at the root node.

A.35.17.3.1.2 Value Type

Value Type (0040,A040) in Content Sequence (0040,A730) of the SR Document Content Module is constrained to the following Enumerated Values (see Table C.17.3-7 for Value Type definitions):

Enumerated Values:

TEXT

CODE

NUM

DATETIME

UIDREF

PNAME

CONTAINER

IMAGE

SCOORD

WAVEFORM

TCOORD

A.35.17.3.1.3 Relationship Constraints

Relationships between Content Items in the content of this IOD shall be conveyed in the by-value mode. See Table C.17.3-8 for Relationship Type definitions.

Note

Relationships by-reference are forbidden. Therefore, Referenced Content Item Identifier (0040,DB73) is not present in any of the Content Items within the SR Document Content Module.

Table A.35.17-2 specifies the relationship constraints of this IOD.

Table A.35.17-2. Relationship Content Constraints for Simplified Adult Echo SR IOD

Source Value Type

Relationship Type (Enumerated Values)

Target Value Type

CONTAINER

CONTAINS

TEXT, CODE, NUM, DATETIME, UIDREF, PNAME, CONTAINER

TEXT, CODE, NUM

HAS OBS CONTEXT

TEXT, CODE, NUM, DATETIME, UIDREF, PNAME, CONTAINER

CONTAINER

HAS ACQ CONTEXT

TEXT, CODE, NUM, DATETIME, UIDREF, PNAME, CONTAINER

any type

HAS CONCEPT MOD

CODE, TEXT

TEXT, CODE, NUM

HAS PROPERTIES

TEXT, CODE, NUM, DATETIME, UIDREF, PNAME, CONTAINER

TEXT, CODE, NUM

INFERRED FROM

TEXT, CODE, NUM, DATETIME, UIDREF, CONTAINER, IMAGE, SCOORD, WAVEFORM, TCOORD

SCOORD

SELECTED FROM

IMAGE

TCOORD

SELECTED FROM

WAVEFORM


DICOM PS3.3 2020b - Information Object Definitions