DICOM PS3.3 2019c - Information Object Definitions

A.35.14.3 Radiopharmaceutical Radiation Dose SR IOD Module Table

Table A.35.14-1 specifies the Modules of the Radiopharmaceutical Radiation Dose SR IOD.

Table A.35.14-1. Radiopharmaceutical Radiation Dose SR IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

SR Document Series

C.17.1

M

Clinical Trial Series

C.7.3.2

U

Frame of Reference

Synchronization

C.7.4.2

C - shall be present if system time is synchronized to an external reference. May be present otherwise.

Equipment

General Equipment

C.7.5.1

M

Enhanced General Equipment

C.7.5.2

M

Document

SR Document General

C.17.2

M

SR Document Content

C.17.3

M

SOP Common

C.12.1

M


A.35.14.3.1 Radiopharmaceutical Radiation Dose SR IOD Content Constraints
A.35.14.3.1.1 Template

The document may be constructed from Baseline TID 10021 “Radiopharmaceutical Radiation Dose” invoked at the root node.

A.35.14.3.1.2 Value Type

Value Type (0040,A040) in Content Sequence (0040,A730) of the SR Document Content Module is constrained to the following Enumerated Values (see Table C.17.3-7 for Value Type definitions):

Enumerated Values:

TEXT

CODE

NUM

DATETIME

UIDREF

PNAME

CONTAINER

A.35.14.3.1.3 Relationship Constraints

Relationships between Content Items in the content of this IOD shall be conveyed by-value. Table A.35.14-2 specifies the relationship constraints of this IOD. See Table C.17.3-8 for Relationship Type definitions.

Table A.35.14-2. Relationship Content Constraints for Radiopharmaceutical Radiation Dose SR IOD

Source Value Type

Relationship Type (Enumerated Values)

Target Value Type

CONTAINER

CONTAINS

TEXT, CODE, NUM, DATETIME, UIDREF, PNAME, CONTAINER

TEXT, CODE, NUM

HAS OBS CONTEXT

TEXT, CODE, NUM, DATETIME, UIDREF, PNAME

CONTAINER

HAS ACQ CONTEXT

TEXT, CODE, NUM, DATETIME, UIDREF, PNAME, CONTAINER.

any type

HAS CONCEPT MOD

TEXT, CODE

TEXT, CODE, NUM, PNAME

HAS PROPERTIES

TEXT, CODE, NUM, DATETIME, UIDREF, PNAME, CONTAINER.

TEXT, CODE, NUM

INFERRED FROM

TEXT, CODE, NUM, DATETIME, UIDREF, CONTAINER.


DICOM PS3.3 2019c - Information Object Definitions