DICOM PS3.3 2021d - Information Object Definitions

10.23 RT Equipment Correlation Macro

Table 10-27. RT Equipment Correlation Macro Attributes Description

Attribute Name

Tag

Type

Attribute Description

Patient Support Angle

(300A,0122)

3

Patient Support angle, i.e., orientation of IEC PATIENT SUPPORT coordinate system with respect to IEC FIXED REFERENCE coordinate system (degrees).

Table Top Pitch Angle

(300A,0140)

3

Table Top Pitch Angle, i.e., the rotation of the IEC TABLE TOP coordinate system about the X-axis of the IEC TABLE TOP coordinate system (degrees). See Section C.8.8.25.6.2.

Table Top Roll Angle

(300A,0144)

3

Table Top Roll Angle, i.e., the rotation of the IEC TABLE TOP coordinate system about the Y-axis of the IEC TABLE TOP coordinate system (degrees). See Section C.8.8.25.6.2.

Table Top Longitudinal Position

(300A,0129)

3

Table Top Longitudinal position in IEC TABLE TOP coordinate system (mm).

Table Top Lateral Position

(300A,012A)

3

Table Top Lateral position in IEC TABLE TOP coordinate system (mm).


DICOM PS3.3 2021d - Information Object Definitions