DICOM PS3.2 2021d - Conformance

L.4.2 AE Specifications

This AE complies with Section 6.2 “Transport” in PS3.22 , specification for DICOM-RTV.

L.4.2.1 DICOM-RTV Application Entity Specifications

L.4.2.1.1 SOP Classes

EXAMPLE-RTV-SERVICE provides Standard Conformance to the following SOP Classes:

Table L.4.2-1. SOP Classes for DICOM-RTV AE

SOP Class Name

SOP Class UID

SCU

SCP

Video Photographic Image Real-Time Communication

1.2.840.10008.10.2

No

Yes


Some restrictions applies on the Real-Time Communications:

Table L.4.2-2. DICOM-RTV Instances Specification

Category

Restrictions

Transfer Syntaxes Supported

SMPTE ST 2110-20 Uncompressed Progressive Active Video

Photometric interpretation

RGB

Bit depth

10


Table L.4.2-3. DICOM-RTV Screen Resolutions

Rows

Columns

Frame rate

Video Type

Progressive or Interlaced

1080

1920

25

25 Hz HD

P

1080

1920

29.97, 30

30 Hz HD

P

1080

1920

25

25 Hz HD

I

1080

1920

29.97, 30

30 Hz HD

I

720

1280

25

25 Hz HD

P

720

1280

29.97, 30

30 Hz HD

P

720

1280

50

50 Hz HD

P

720

1280

59.94, 60

60 Hz HD

P


The resolution is defined by the equipment configuration, and is reflected in the SDP object.

L.4.2.1.2 Connection Policies
L.4.2.1.2.1 General

The consumer shall get the SDP object on the following URL: http://<local-IP-address-of-the-device>/SDP.

L.4.2.1.2.2 Number of Connections

L.4.2.1.2.2

EXAMPLE-RTV-SERVICE is provided in multicast. The limit of simultaneous connection depends on the local network infrastructure.

DICOM PS3.2 2021d - Conformance